40
KF format
5if type 2
This is the Karl-Fischer moisture meter format and is used when the peripheral equipment can only
communicate using this format.
This format consists of fourteen characters excluding the terminator.
This format has no header characters.
The polarity sign is placed before the data, with spaces in place of leading zeros, if the data is
not zero or overloaded.
This format outputs the unit only for a stable value.
MT format
5if type 3
A header of two characters indicates the balance condition.
The polarity sign is used only for negative data.
The weighing data uses spaces in place of the leading zeros.
The character length of this format changes dependent upon the unit
NU (Numerical) format
5if type 4
This format outputs only numerical data.
This format consists of nine or ten characters excluding the terminator.
With the MC-1000, when the numerical data excluding the decimal point exceeds eight
characters, the data format will be ten characters long.
The polarity sign is placed before the data with the leading zeros. If the data is zero, the plus
sign is used.
(Data example exceeding 8 characters)
+ 1 0 0 0 . 0 0 0 0 g
CR LF
Data
Unit
Terminator
S
1 0 0 0 . 0 0 0 0 g
CR LF
Header
Unit Terminator
Data
+ 0 0 0 . 0 1 2 7
CR LF
Data
Terminator
+ 1 0 0 0 . 0 0 0 0
CR LF
Data
Terminator
g
Stable value
Unstable value
Summary of Contents for MC-1000
Page 1: ...MC 1000 MC 6100 1WMPD4002204B...
Page 7: ...5...
Page 68: ...66 20 EXTERNAL DIMENSIONS MC 1000 MC 6100 Unit mm...
Page 73: ...71 MEMO...
Page 74: ...72 MEMO...
Page 75: ...13...