![A&D HV-C Series Instruction Manual Download Page 27](http://html1.mh-extra.com/html/aandd/hv-c-series/hv-c-series_instruction-manual_3944515027.webp)
HV/HW-C/CP Series
Page 25
7.1.2. When Using the Batteries for Type C
Step 1 Install four new batteries. Refer to "
5.2. Installing the Batteries for Type C
".
Step 2 Confirm that nothing is placed on the weighing pan.
Step 3 The scale is turned on or off using the ON/OFF key.
Step 4 Check the accuracy of weighing. Allow a 30-minute warm up period before calibration.
Step 5 With nothing on the weighing pan, press the ZERO key to display zero value.
Step 6 Gently place an article to be weighed on the weighing pan.
Step 7 Wait for the stability mark to be displayed. Read the weighing value.
Step 8 Gently remove the article from the weighing pan.
Step 9 Turn the scale off using the ON/OFF key.
Caution
When
is displayed, it means that the batteries will run out soon.
Replace used batteries with four new ones when
is displayed.
Battery life depends on the ambient temperature.
Remove the batteries from the display unit when the scale is not to be used for a long time.
Leaving them installed may result in leakage and cause a malfunction and damage to the scale.
7.1.3. Power on Zero Function and Power on Tare Function
Power on Zero Function :
When nothing is placed on the weighing pan and the scale is turned on using the ON/OFF
key, the weighing value is assumed as the reference value of weighing. The zero value as
gross value and
zero point mark
are displayed. The
power on zero
function can be used
when the weighing value is within the threshold value
from the zero point at calibration.
Power on Tare Function :
When a tare weight (container) is placed on the weighing pan and the scale is turned on using
the
ON/OFF
key, the tare function is performed and a zero value is displayed as net value.
Zero point mark
and
net mark
are displayed. The
power on tare
function can be used when
the current zero point is more than the threshold value
from the zero point of calibration.
Threshold value
Active function
Principal standard models
Approved models
Power on zero
function
±
50
% of capacity
±
10
% of capacity
Power on tare
function
+
50
% of capacity
+
10
% of capacity