Automatic power-off function

The automatic power-off function turns power off after approximately 2
minutes of stable display during the object(s) applied. When there is no
object applied, it will turn off after 30 seconds.

Calibration

1. Make sure that there is nothing on the weighing pan, press and hold

the  LIGHT/CAL   key approximately 3 seconds. The display shows

 Cal 

. Release the key.

2. The scale memorizes the zero point and automatically displays

100.0

g .

3. Place the calibration weight of 100g on the center of the pan within 10

seconds.

4. After approximately 8 seconds, the scale will display     

f

 for

success and     

e

  for failure. Then, the scale automatically turns off.

5. Turn the scale on again to start the operation.

G

 If calibration fails, check if anything is touching the weighing pan,

there is vibration or wind, and the calibration weight is correct.

Dimension

(mm)

                  Scale                                      Protective cover

60

20.

3

64

98

66

70

10

2

Summary of Contents for HJ-150

Page 1: ...iod of time remove the batteries G Avoid over shock load that could cause damage to the scale G Keep the protective cover attached when you carry the scale G Wait for more than 6 seconds to let the scale warm up after turning on the power G When the location to be used or the ambient environment has changed calibrate the scale before use Specifications Capacity 150 x 0 1 g 5 29 x 0 01 oz Display 7...

Page 2: ... when installing and or replacing batteries to prevent excessive force from being added to the weighing pan Weighing procedures 1 Turn on off the scale Press the ON ZERO OFF key to turn on the scale The scale shows all segments for a few seconds and displays zero The backlight will turn on for about one minute Press and hold approximately 3 seconds to turn the scale off 2 Select the weighing unit ...

Page 3: ...play shows Cal Release the key 2 The scale memorizes the zero point and automatically displays 100 0 g 3 Place the calibration weight of 100g on the center of the pan within 10 seconds 4 After approximately 8 seconds the scale will display f for success and e for failure Then the scale automatically turns off 5 Turn the scale on again to start the operation G If calibration fails check if anything...

Page 4: ... Abingdon Business Park Abingdon Oxon OX14 1DY U K Telephone 44 1235 550420 Fax 44 1235 550485 German Sales Office Große Straße 13b 22926 Ahrensburg GERMANY Telephone 49 0 4102 459230 Fax 49 0 4102 459231 A D MERCURY PTY LTD 32 Dew Street Thebarton South Australia 5031 AUSTRALIA Telephone 61 8 8301 8100 Fax 61 8 8352 7409 A D KOREA Limited Manhattan Bldg 8F 36 2 Yoido dong Youngdeungpo gu Seoul KO...

Reviews: