This manual describes how the GXA-25 External fanless ionizer (static eliminator) work and how to 

get the most out of them in terms of performance. Read this manual thoroughly before using the 

ionizer and keep it at hand for future reference. 

 

1.

Features of the Option 

□ 

A fanless DC ionizer (static eliminator) can eliminate static electricity from the weighing sample 
before the measurement, reducing weighing error. 
(Bipolar ions are generated by DC corona discharge and the target is irradiated by those to 
neutralize it.) 

□ 

Each electrode unit of the ionizer is designed to be removed, cleaned and replaced. 

□ 

An IR (touchless) switch is attached to the ionizer, and neutralization can be started without 
touching the ionizer. 

□ 

Power is supplied from the balance to allow the ionizer to be operated without using an AC 
adapter. 

□ 

PRINT or RE-ZERO or the neutralization function for the ionizer can be operated by using the 
accessory foot switch (AX-SW137-PRINT or AX-SW137-REZERO).

 

 

Static electricity 
In general, when the ambient humidity is less than 45%RH, powders, paper, plastic, nonconductors, 
etc., easily become charged with static electricity. The influence of the static electricity may cause a 
weighing error of several milligrams. The ionizer effectively neutralizes the electrical charge. 
 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Summary of Contents for GXA-25

Page 1: ...GXA 25 External fanless ionizer 1WMPD4003577...

Page 2: ...duced transmitted transcribed or translated into any language in any form by any means without the written permission ofA D Company Ltd The contents of this manual and the specifications of the instru...

Page 3: ...ames 3 1 2 How to Install 3 1 3 How to Use 4 1 5 Controlling the Ionizer Externally 6 2 Function Settings of the Balance 7 2 1 Display for Function Settings and Operation Keys 7 2 2 Items List 8 3 Mai...

Page 4: ...de unit of the ionizer is designed to be removed cleaned and replaced An IR touchless switch is attached to the ionizer and neutralization can be started without touching the ionizer Power is supplied...

Page 5: ...the panel with double sided tape and insert it into the socket of the option board 3 Attach the option board to the balance Secure with the two screws removed in step 1 Balance rear side 4 Connect usi...

Page 6: ...charged The electrodes must not be touched with tweezers etc while neutralizing 1 Place the weighing sample to be neutralized in front of the ionizer At this time the weighing sample to be neutralize...

Page 7: ...is operated and then the balance automatically returns to the previous weighing mode when operations for the ionizer are completed While neutralizing make sure that the weighing sample is lifted up a...

Page 8: ...ated to stop neutralization When the AX SW137 PRINT is connected to the balance Class Item Parameter Details and usage 0 ION key Assigns a key for starting neutralization to the AX SW137 PRINT The IR...

Page 9: ...s and use conditions use environment Carefully confirm the contents to be changed before changing them 2 1 Display for function settings and operation keys The indicates that the parameter displayed i...

Page 10: ...ssword locking auto Cal Auto calibration Refer to balance s instruction manual 0 1 second 1 3 seconds 2 10 seconds ion Neutralization interval 3 Manual for up to 10 minutes Only external control is en...

Page 11: ...ctrodes while the mark and the LED are flashing 4 Specifications Static elimination method DC corona discharge Bipolar Elimination range Approx 10 cm to 30 cm from the electrode Elimination performanc...

Page 12: ...MEMO...

Page 13: ...MEMO...

Page 14: ...MEMO...

Page 15: ......

Page 16: ...1235 550485 A D AUSTRALASIA PTY LTD 32 Dew Street Thebarton South Australia 5031 AUSTRALIA Telephone 61 8 8301 8100 Fax 61 8 8352 7409 A D KOREA Limited 6 33 817 150 749 817 Manhattan Bldg 33 Gukjege...

Reviews: