EN
8
4-2. Calibration Using an External Weight
This function calibrates the balance using an external weight.
1. Connect the AC adapter and warm up the balance for at
least 30 minutes with nothing on the weighing pan.
2. Press and hold the
key for 2 seconds until
is displayed, and then release the key.
3. The balance displays
.
If you want to change the calibration weight, press
the
key and go to step 4.
Refer to "8. CALIBRATION" in the instruction manual
for selectable calibration weights.
If you use the calibration weight value stored in the
balance, go to step 5.
4. Specify the calibration weight value as follows:
key To switch the display condition to "All
of the segments blinking" (target
weight selection mode) or "The last
two digits blinking" (value adjustment
mode).
key (To increase the value)
key (To decrease the value)
To select the calibration weight or
adjust the value.
key
To store a new weight value. Even if the
AC adapter is removed, the data is
maintained in non-volatile memory.
key
To cancel the operation (return to
).
5. Confirm that there is nothing on the pan and press the
key. The balance measures the zero point and
displays the measured value. Do not allow vibration or
drafts to affect the balance.
6. Place the calibration weight on the pan and press the
key. The balance measures the calibration
weight and displays the measured value. Do not allow
vibration or drafts to affect the balance.
7. Remove the weight from the pan.
8. After the calibration, if the “GLP output” is set, the
balance outputs a “sensivity adjustment report”.
9. The balance will automatically return to the weighing
mode.
10. Place the calibration weight on the pan again and check if
the value is within the range (set value ± 2 digits). If it is
out of the range, pay attention to the surrounding
environment and start again from step 1.
Press and hold the
key for 2 seconds
Release the key
Press the
key
Select a
weight
e.g. Calibration weight
value: 1000.12g
Calibration weight
GLP output
Summary of Contents for FZ-i Series
Page 54: ...CT 54...
Page 55: ...CT 55 3 2 2 ON OFF ON OFF FZ i FZ i WP FZ GD GLP GMP Title block End block pcs ct mom g 2...
Page 56: ...CT 56 4 4 1 FZ i FZ iWP FZ GD 1 30 2 3 4 GLP output GLP 5...
Page 58: ...CT 58 5 EC E11 3 2 During Use 3 2 EC E17 FZ i FZ i WP FZ GD EC E20 EC E21 A D EC E00 EC E01...
Page 59: ...CT 59 EC E02 Q RE ZERO Q EC E03 1 EC E04 EC E06 EC E07...
Page 61: ......