AX
-
ION
-
25
Ionizer
INSTRUCTION MANUAL
2020 A&D Company Ltd. All rights reserved.
No part of this publication may be reproduced, transmitted,
transcribed, or translated into any language in any form by
any means without the written permission of A&D
Company Ltd.
The contents of this manual and the specifications of the
instrument covered by this manual are subject to change
for improvement without notice.
1WMPD4004253A
1.
Introduction
This manual describes outline of the ionizer (AX
-
ION
-
25)
and how to use it correctly. Please thoroughly read this
manual before using the AX
-
ION
-
25 and keep it at hand for
future reference.
2.
Features
This ionizer eliminate static electricity on the weighing sample by
generating bipolar ions from 4 electrode needles.
The ionizer can eliminate static electricity on a charged
weighing sample without breeze. A weighing error can be
reduced.
AC adaptor is needless because power of the ionizer is
supplied from main unit of the balance.
This ionizer can be controlled by using built-in IR sensor,
option IR switch (AX
-
IR
-
SWITCH).
Electrode unit is removable. A unit can be cleaned and
replaced.
Static electricity
In general, nonconductors such as powder, filter,
medicine wrapping paper, plastic etc. easily become
electrostatically charged when the ambient humidity is
less than 45 %RH. The static electricity may cause
weighing error of approximately few mg at weighing.
This ionizer can perform static elimination effectively.
3.
Cautions before Use
Read the following cautions for safety use of the ionizer.
Operate the ionizer gently because of precision instrument.
Do not install the following place.
Place getting water, vibration, shock, direct sunshine. Dusty place, air
including salt or corrosive gas, a place in flammable gas.
Do not turn on power of the ionizer until the installation is finished. The
switch to turn off is not equipped in the ionizer.
4.
Specifications
Ion generation method
Direct corona discharge
Effective range of static
electricity elimination
Approximately 7 cm or shorter from the
electrode needle
Operation temperature and
humidity
5 to 40 °C,
85% RH or less (no condensation)
Discharge electrode needle Tungsten (with a life of approx. 10,000 hours)
Dimensions 68(W)×128(D)×163(H)
mm
Mass Approximately
370
g
5.
Part Names
Discharge electrode unit
Front
IR SENSOR
Touchless infrared
proximity sensor
Power lamp
Lights when turning on.
ION lamp
Lights when neutralizing.
Rear Optimizing
switch
Connecter for option
IR switch
Connector for
connecting balance
Connection cable (60 cm)
6.
How to Use
6.1.
Connecting to the balance
Caution
Disconnect the AC adapter before
installing the ionizer to the balance.
1. Connect the ionizer to the balance using
connection cable.
2. Connecting the AC adapter of the
balance, the power lamp lights up.
The ionizer supports only BA series as of
May 2021. Refer the following homepage
for the updated information.
https://www.aandd.jp
Balance
Rear
Connection
cable
(60
cm)
6.2.
Static elimination
1. Put the weighing sample into the effective range shown in the figure 1.
2. Responding the IR sensor (touchless infrared proximity sensor) on the front
panel, option IR switch, the ION lamp lights up and static elimination starts.
3. Perform static elimination referring to the figure 1. Static elimination process
stops automatically in 3 seconds and ION lamp turns off for factory setting.
Referring to "
7. Optimizing the Ionizer
", optimize static elimination method
and discharging time if the distance between electrode needles and
weighing sample is 10 cm or father.
Top view
Effective
range
0
cm
5
cm
7
cm
10
cm
Range
Side view
Effective
range
1
sec.
1.5
sec.
3
sec.
Discharge
time
Figure 1 Effective range of static elimination
6.3.
Connecting Options
Caution
Remove power supply cable when connecting peripherals to
the ionizer.
Static elimination process can be controlled by using
option IR switch instead of the built-in IR sensor. The
ionizer can not send PRINT and RE-ZERO command
to the balance by using option IR switch
Option IR switch
7. Optimizing the Ionizer
Static elimination method can be optimized by using switches on the
rear. Refer to the following table for functions of switches. Change the
settings of the switch 1 and switch 2 eliminating static electricity
outside the effective range shown as in the figure 1.
ON
ON
Factory settings of switches is all ON.
Knob of switch
OFF
1 2 3 4
Switch
1 2 3 4
Factory settings of switches is all ON (Upper side). It means static
elimination method is "Timer mode", "Discharging time" is 3 seconds,
"Built-in IR sensor" and "Buzzer" is available.
Table 1 Functions of the optimizing switches
Switch No.
/
Item
State of switch
Descriptions
Switch 1
Static elimination method
ON Upper side Timer mode
#1
OFF Lower side Manual mode
#2
Switch 2
Discharging
time
#3
ON Upper side 3 seconds
OFF Lower side 10 seconds
Switch 3
Built
-
in IR sensor
ON Upper side Available
OFF Lower side Not Available
Switch 4
Buzzer
#4
ON Upper side Available
OFF Lower side Not Available
#1
When switch 1 is selected "Timer mode", static elimination is
performed for "Discharging time" set by switch 2.
#2
Once static elimination has started, it continues until built-in IR
sensor or option IR switch responds again when switch 1 is selected
"Manual mode". "Manual mode" does not stop static
elimination automatically.
#3
The setting of the switch 2 is effective when switch 1 is
selected "Timer mode".
#4
Buzzer sounds at power on and responding the built-in
IR sensor or option IR switch.
8.
Maintenance
Caution
Do not touch the discharge electrode unit to avoid
electric shock during neutralization.
When the ionizer is used continuously, discharge
electrode needle and around may get dirty and
neutralization performance may become weak. Clean the
electrode needle periodically using cotton swabs to
maintain performance.
When the electrode needle wears out and static elimination
performance does not refresh by cleaning, replace all of
discharge electrode units to new ones of option. Life time of
discharge electrode unit is approximately 10000 hours.
Procedure of replacement
1. Remove the connection cable to turn off the ionizer.
2. Rotate a discharge electrode unit to 45 degree
counterclockwise. Remove it.
3. Insert new units and rotate them to 45 degree clockwise.
Needle
Clean
Step 2
Step 3
Discharge
electrode
unit
9.
Options
Discharge electrode unit
AX
-
BM
-
NEEDLESET (A set is 4 units.)
Replacement electrode units.
Replace all of 4 units with new ones.
Refer to the "
8. Maintenance
" for
replacement procedure.
IR switch
AX
-
IR
-
SWITCH
Touchless infrared proximity sensor.
Refer to the "
6.3. Connecting
Options
" how to connect it.
Static elimination operation can be
controlled by moving hand over the
"SENSOR".