1.

 

Introduction 

This manual describes outline of the ionizer (AX

-

ION

-

25) 

and how to use it correctly. Please thoroughly read this 
manual before using the AX

-

ION

-

25 and keep it at hand for 

future reference.  

 

2.

 

Features 

This ionizer eliminate static electricity on the weighing sample by 
generating bipolar ions from 4 electrode needles.  

  The ionizer can eliminate static electricity on a charged 

weighing sample without breeze. A weighing error can be 
reduced.  

  AC adaptor is needless because power of the ionizer is 

supplied from main unit of the balance.  

  This ionizer can be controlled by using built-in IR sensor, 

option IR switch (AX

-

IR

-

SWITCH).  

  Electrode unit is removable. A unit can be cleaned and 

replaced.  

Static electricity

 

In general, nonconductors such as powder, filter, 
medicine wrapping paper, plastic etc. easily become 
electrostatically charged when the ambient humidity is 
less than 45 %RH. The static electricity may cause 
weighing error of approximately few mg at weighing.  
This ionizer can perform static elimination effectively. 

 

3.

 

Cautions before Use 

Read the following cautions for safety use of the ionizer.  

  Operate the ionizer gently because of precision instrument.  
  Do not install the following place.  

Place getting water, vibration, shock, direct sunshine. Dusty place, air 
including salt or corrosive gas, a place in flammable gas. 

  Do not turn on power of the ionizer until the installation is finished. The 

switch to turn off is not equipped in the ionizer.  

 

4.

 

Specifications 

 

Ion generation method 

Direct corona discharge 

Effective range of static 

electricity elimination 

Approximately 7 cm or shorter from the 

electrode needle  

Operation temperature and 
humidity 

5 to 40 °C,  
85% RH or less (no condensation)

Discharge electrode needle  Tungsten (with a life of approx. 10,000 hours) 
Dimensions 68(W)×128(D)×163(H) 

mm 

Mass Approximately 

370 

 

5.

 

Part Names 

 
Discharge electrode unit 

Front 

 
IR SENSOR 
Touchless infrared 
proximity sensor 
 
Power lamp 
Lights when turning on. 
 
ION lamp 
Lights when neutralizing. 

 Rear  Optimizing 

switch 

 

Connecter for option  
IR switch 
 
Connector for  
connecting balance 
 
 

 

Connection cable (60 cm) 

 
 

6.

 

How to Use 

6.1.

 

Connecting to the balance 

Caution

  Disconnect the AC adapter before 

installing the ionizer to the balance. 

1.  Connect the ionizer to the balance using 

connection cable.  

2.  Connecting the AC adapter of the 

balance, the power lamp lights up.  

 The ionizer supports only BA series as of 

May 2021. Refer the following homepage 
for the updated information. 

https://www.aandd.jp 

 

 
 
 Balance 
 

 Rear 
 
 
 Connection 

cable 

(60 

cm) 

 

6.2.

 

Static elimination 

1.  Put the weighing sample into the effective range shown in the figure 1. 
2.  Responding the IR sensor (touchless infrared proximity sensor) on the front 

panel, option IR switch, the ION lamp lights up and static elimination starts.  

3.  Perform static elimination referring to the figure 1. Static elimination process 

stops automatically in 3 seconds and ION lamp turns off for factory setting. 
Referring to "

7. Optimizing the Ionizer

", optimize static elimination method 

and discharging time if the distance between electrode needles and 
weighing sample is 10 cm or father.  

 

 

Top view 

 
 Effective 

range 

 

 

 

cm 

5

 cm 

7

 cm 

10

 cm 

 Range 
Side view 

 

 Effective 

range 

 
 

 

 

1

 sec. 

1.5

 sec. 

3

 sec. 

 Discharge 

time 

 

 

Figure 1  Effective range of static elimination

 

 

6.3.

 

Connecting Options 

Caution

  Remove power supply cable when connecting peripherals to 

the ionizer.  

 

Static elimination process can be controlled by using 
option IR switch instead of the built-in IR sensor. The 
ionizer can not send PRINT and RE-ZERO command 
to the balance by using option IR switch  
 
 
Option IR switch 
 
 

 

7. Optimizing the Ionizer 

Static elimination method can be optimized by using switches on the 
rear. Refer to the following table for functions of switches. Change the 
settings of the switch 1 and switch 2 eliminating static electricity 
outside the effective range shown as in the figure 1.  
 ON 

ON 

Factory settings of switches is all ON. 

 

Knob of switch 

 

OFF 

1 2 3 4   

 Switch 

1 2  3  4 

Factory settings of switches is all ON (Upper side). It means static 
elimination method is "Timer mode", "Discharging time" is 3 seconds, 
"Built-in IR sensor" and "Buzzer" is available.  

Table 1  Functions of the optimizing switches 

Switch No. 

/

 Item 

State of switch 

Descriptions 

Switch 1 
 

Static elimination method 

 ON  Upper side Timer mode 

#1

 

 OFF  Lower side Manual mode 

#2

 

Switch 2 
 Discharging 

time 

#3

 

 ON  Upper side 3 seconds 
 OFF  Lower side 10 seconds 

Switch 3 
 Built

-

in IR sensor 

 ON  Upper side Available 
 OFF  Lower side Not Available 

Switch 4 
 Buzzer 

#4

 

 ON  Upper side Available 
 OFF  Lower side Not Available 

#1

  When switch 1 is selected "Timer mode", static elimination is 

performed for "Discharging time" set by switch 2.  

#2

  Once static elimination has started, it continues until built-in IR 

sensor or option IR switch responds again when switch 1 is selected 

"Manual mode". "Manual mode" does not stop static 
elimination automatically.  

#3

  The setting of the switch 2 is effective when switch 1 is 

selected "Timer mode".  

#4

  Buzzer sounds at power on and responding the built-in 

IR sensor or option IR switch.  

 

8.

 

Maintenance 

Caution

  Do not touch the discharge electrode unit to avoid 

electric shock during neutralization.  

  When the ionizer is used continuously, discharge 

electrode needle and around may get dirty and 
neutralization performance may become weak. Clean the 
electrode needle periodically using cotton swabs to 
maintain performance.  

  When the electrode needle wears out and static elimination 

performance does not refresh by cleaning, replace all of 
discharge electrode units to new ones of option. Life time of 
discharge electrode unit is approximately 10000 hours.  

Procedure of replacement 

1.  Remove the connection cable to turn off the ionizer.  
2.  Rotate a discharge electrode unit to 45 degree 

counterclockwise. Remove it.  

3.  Insert new units and rotate them to 45 degree clockwise.  

 Needle 

Clean 

 

Step 2 

Step 3 

 
 
 
 
 

 Discharge 

electrode 

unit 

 

9.

 

Options 

Discharge electrode unit

 AX

-

BM

-

NEEDLESET (A set is 4 units.) 

 Replacement electrode units.  
  Replace all of 4 units with new ones. 
  Refer to the "

8. Maintenance

" for 

replacement procedure.  

 

IR switch

 AX

-

IR

-

SWITCH 

 Touchless infrared proximity sensor.  
  Refer to the "

6.3. Connecting 

Options

" how to connect it.  

  Static elimination operation can be 

controlled by moving hand over the 
"SENSOR".  

 

 