6.
Functions
There are two parameters lists, one for the F-functions and one for the CF-functions.
These functions control the indicator. The parameters of each function are stored in
the non-volatile memory, and are not lost even if power is turned off.
F-functions: These parameters can always be changed and are used for internal settings.
CF-functions: If you accept a certified approval of the weighing instruments, the CAL
cover (rear panel) must be sealed. In this case, parameters of the
CF-function can not be changed.
6.1.
Changing the Function Settings
To enter the function setting mode, do either of the following.
1. When the display is off, press the
ON/OFF
key while pressing the
+/-
key.
2. When in the weighing mode, press both the
+/-
and
ON/OFF
keys at the same time.
In the function setting mode,
f00
will be displayed .
Selecting an Item
Step 1 Select an item using the
0
-
9
keys.
ENTER
key The key to display a parameter
of the selected item.
Proceed to step 2.
ESC
key
To finish the function setting mode
and enter the weighing mode.
CAL
key
The key to switch between
F-functions and CF-functions.
Selecting a Parameter
Step 2 Select a parameter using the
0
-
9
keys.
ENTER
key The key to store the parameter
and return to step 1.
ESC
key
The key to return to step 1
without changing the parameter.
CLEAR
key The key to restore the display
to the initial setting.(Type1)
Changes the sub item.(Type2,
Type3) Type2 and Type3 are
indicated in the parameter table.
Exiting from the Function Setting Mode
Press
ESC
key to proceed to the weighing mode when an item number is
displayed.
5.6. Calibration Error Code List
6. Functions
Weighing
Weighing
+/-
NET/GROSS
ENTER
ENTER
ENTER
ENTER
TARE
ESC
9
0
9
0
+/-
NET/GROSS
ENTER