background image

 

2

2. Specifications 

AD-4329A-07 Analog 

output 

Current output 

4mA to 20mA 

 

Outputs 4mA or less when the display is turned off or the indicator 

is in a mode other than the weighing data displaying mode, such 

as calibration. 

 

Outputs 20mA or more at overload display. 

Non-linearity ±0.1%FS 

max. 

Load resistance 

0

 to 520

 

Zero temperature coefficient  ±0.02 %FS/°C max. 

Span temperature coefficient  ±0.02 %FS/°C max. 

Resolution Approx. 

1/3000 

Output adjustment 

SPAN control and ZERO control located on the option board 

Terminal 

M3 compression terminal 

Note: Analog output can not be used at the same time as RS-232C output or other options 

except OP-02. 

 

Pin Connection 

 

 

 

 

 

 

 

Related settings 

Settings related to analog output are performed by the function settings, 

f-60

,

 f-61

, and 

f-62

.   

For details on the setting procedure, refer to the AD-4329A instruction manual. 

Item Parameter 

Description 

f-60

  

0

 

Displayed value 

f-60

  

1

 

Gross data 

Output data 

f-60

  

2

 

Net data 

Value at 4mA output 

f-61

 

 

-999999 to 999999 (Factory setting: 0) 
The decimal point position depends on the 

Cf-00

 

setting. 

Value at 20mA output 

f-62

 

-999999 to 999999 (Factory setting: 10000) 
The decimal point position depends on the 

Cf-00

 

setting. 

Note: To confirm the F-61 and F-62 settings, press the                key and preceed to the next item. 

Do not press the MODE key. 

 

Reviews: