SAFETY PRECAUTIONS

 

All safety messages are identified by the following, "

WARNING

" or 

"

CAUTION

", of ANSI Z535.4 

(American National Standard Institute: Product Safety 

Signs and Labels)

. The meanings are as follows: 

 
 
 
 
 



This manual is subject to change without notice at any time to improve the product. 



Product specifications are subject to change without any obligation on 
the part of the manufacturer. 



When using the AD-1611, the following safety precautions should 
always be observed. 

 

CAUTION 

 



Do not disassemble the AD-1611. Disassembling may cause damage to 
the AD-1611. Damage caused by disassembling will not be covered by 
the warranty. Contact your local A&D dealer if the AD-1611 needs 
service or repair.   

 



If a problem has occurred and you cannot resolve it, stop using the 
AD-1611.  

 
 

PRECAUTIONS FOR USE

 

AD-1611 main body 



The AD-1611 is not explosion-proof or waterproof. Do not place it where 
there is a risk of explosion or where it may get wet.   



When the optical port is not used or the AD-1611 is stored, attach the 
protective cap provided to the port. Otherwise, dust may enter the 
connector in the optical port and cause a communication error. 

 

Power supply 



Power is supplied to the AD-1611 by the accessory AC adapter or by a 
personal computer (PC) connected using the accessory USB cable. 



When the USB cable is used for communication, power is supplied by a 
PC. Only a PC can be connected to the USB port. Do not connect other 
devices to the USB port nor take power from an outside source. 
When an RS-232C cable is used for communication, use the AC 
adapter to supply power. 

WARNING

 

A potentially hazardous situation which, if not   
avoided, could result in death or serious injury.

 

CAUTION

 

A potentially hazardous situation which, if not   
avoided, may result in minor or moderate injury.

 

Summary of Contents for AD-1611

Page 1: ...1WMPD4003539...

Page 2: ...FOR USE AD 1611 main body The AD 1611 is not explosion proof or waterproof Do not place it where there is a risk of explosion or where it may get wet When the optical port is not used or the AD 1611...

Page 3: ...terminal to ground the AD 1611 Refer to 2 1 AD 1611 Optical fiber cable Use the recommended cables below Using other cables may cause damage to the connector or may cause a communication error Cables...

Page 4: ...e or a PC When connected to a balance scale with the optical connector the AD 1611 converts optical signals to electric signals and output them through RS 232C or USB 2 PARTS DESCRIPTION ACCESSORIES 2...

Page 5: ...essories are provided as standard and options are sold separately 3 2 Preparation 1 When USB is used Install a USB driver on a PC A USB driver is available for download at our website http www aandd j...

Page 6: ...setting is required for the AD 1611 When the AD 1611 is used for the first time the cables are reconnected or the power is turned on such as after the balance scale battery replacement check as follow...

Page 7: ...1 below Receives data 2 The balance scale PRINT key is not pressed Or The AC adapter or USB is not connected Off Receives no data Is turned off Note 1 Each time the balance scale PRINT key is pressed...

Page 8: ...cale for malfunctioning The optical fiber cable used is not one that is recommended Use a recommended cable 1 The LED does not blink even if the PRINT key is pressed The optical fiber cable is damaged...

Page 9: ...or scales manufactured by A D with a specified optical connector installed Visit our website for available balances or scales Baud rate According to the value set on the balance scale Receiving connec...

Page 10: ...t A D AUSTRALASIA PTY LTD 32 Dew Street Thebarton South Australia 5031 Australia Telephone 61 8 8301 8100 Fax 61 8 8352 7409 http www andaustralasia com au A D KOREA Limited Manhattan Building 8th Flo...