background image

 exposed conductive parts, however, extra pre- 

   caution should be taken to minimize the possibility 

 of 

shorting. 

 

Do not make measurement with the Battery Cover 

removed.

 

 

Disconnect the test leads from the instrument for

current measurement. 

(1) Set the Function Switch to “

600A

” or “

1000A

” position. 

(on DSA680, only “600A” is available) AC has been

selected by default; press the SELECT key, when DC

has been selected, to change it to AC. AC mark is

displayed at the upper left on the display. 

(2) Press the trigger to open the transformer jaws and

 clamp them onto the one conductor under test, then

 

take the reading on the display. Pressing the

 “

Hz/DUTY

” Key switches the indication in following

 sequence. 

AC Current            Hz          DUTY 

Hz/DUTY

 Function requires 50A or more at AC600A 

Range and 350A or more at AC1000A range. 

CAUTION 

 

Max conductor size for DSA680 is approx dia.

33mm and for DSA1020 is approx dia. 40mm. 

During current measurement, keep the trans-

  former jaws fully closed. Other wise, accurate 

 measurements 

cannot 

be 

taken. 

5-2. DC Current Measurement

 

DANGER 

 

Never make measurement on a circuit in which

voltage over DC600V exists to avoid getting

electrical shock. 

 

Do not make measurement with the Battery Cover 

removed.

(1) Set the Function Switch to “

600A

” or “

1000A

” position. 

AC has been selected by default; press the SELECT

key, when AC has been selected, to change it to DC.

(only 600A is available on DSA680) 
DC mark is displayed at the upper left on the display. 

(2) 

With the transformer jaws closed and without

 clamping them onto the conductor, press the “

ZERO

” key 

to zero adjust the display. (

mark is displayed at   the 

upper right on the display.) 

(3) Press the trigger to open the transformer jaws and

clamp them onto the one conductor under test, the

conductor should be at the center of the jaws, then   take 

the reading on the display. 

(4) Set the Function Switch to an appropriate position

 according to current under test. 

(5) Pressing  the  “

ZERO

” key again releases “

ZERO

” 

 function.  (

mark at the upper right on the display

 disappears.) 

5-3. AC Voltage Measurement 

DANGER 

 

Never make measurement on a circuit in which

voltage over AC600V exists to avoid getting

electrical shock. 

 

Do not make measurement with the Battery Cover 

removed. 

 

Keep your fingers behind the barrier on the

instrument during measurement. 

(1) Set the Function Switch to “

ACV

” position. 

(2) Connect the red test lead to V/

 terminal and the

 black test lead to COM terminal. 
(3) Connect the test leads to the circuit under test. Take

the reading on the display. Pressing the “

Hz/DUTY

” key

while reading is indicated on the display switches the

indication in following sequence. 

AC Voltage          Hz          DUTY 

CAUTION

 

Hz/DUTY Function requires AC40V or higher. 

 

To measure a frequency, measure the voltage on 
the electrical circuit in advance. Then press the Hz

/DUTY key to enter into frequency measurement. 

 

Readings of frequency may fluctuate or be

influenced under noisy environment. 

5-4. DC Voltage Measurement 

DANGER 

 

Never make measurement on a circuit in which

 voltage over DC600V exists to avoid getting

electrical shock. 

 

Do not make measurement with the Battery Cover 

removed. 

 

Keep your fingers behind the barrier on the

 instrument during measurement. 

(1) Set the Function Switch to “

DCV

” position. 

(2) Connect the red test lead to V/

 terminal and the 

black test lead to COM terminal. 
(3) Connect the red and black test leads to the positive

(+)and negative (-) sides of the circuit under test

respectively. Take the reading on the display. If the

connection is reversed, the display indicates the “

-

“ mark. 

5-5. Resistance/ Diode/ Cont/ Capacity 

Measurement 

DANGER 

 

Never use the instrument on an energized circuit. 

 

Do not make measurement with the Battery Cover 

removed. 

Resistance 

(1) Set the Function Switch to “

/Diode/Cont/Capacity

” 

position. 
(2) Connect the red test lead to V/

 terminal and the

black test lead to COM terminal. Confirm “

OL

” is

indicated on the display, and then short-circuit the tips of 

test leads to make the indication zero. 

(3) Connect the test leads to the both ends of the resistor 

under test. 

(4) Take the reading on the display. 

CAUTION 

 

Even if short the test lead tips, indicated value may 

not be zero. But this is because of the resistance of 

test leads and not a failure. 

 

When test leads are open, “OL” is indicated on the 

display. 

Continuity 

(1) Set the Function Switch to “

/Diode/Cont/Capacity

” 

position. “

“ has been selected by default; press the

SELECT key to change it to “

Continuity

”  

Resistance        Diode          Cont            Capacity 

(2) Connect the red test lead to V/

 terminal and the

 

black test lead to COM terminal. Confirm “

OL

” is

 indicated on the display and short-circuit the tips of test 

leads.  

Indication should become zero and buzzer sounds. 

(3) Connect the test leads to the both ends of the 

conductor under test. The buzzer sounds, if the 

resistance under test is 100

 or less. 

Diode 

(1) Set the Function Switch to “

/Diode/Cont/Capacity

” 

position. “

 “ has been selected by default; press the

SELECT key to change it to “

Diode

”  

Resistance     Diode 

Cont        Capacity 

(2) Connect the red test lead to V/

 terminal and the

black test lead to COM terminal. 

(3) Connect the red and black test leads to the Anode and 

cathode of the diode under test respectively. 

Take the reading on the display. If the connection is

reversed, the display indicates “

OL

”. 

CAUTION 

 

Some of diodes cannot be tested. Indication on

 the display will be “OL”. 

(Zener diode, LED and so on) 

Capacity 

(1) Set the Function Switch to “

/Diode/Cont/Capacity

” 

position. “

 “ has been selected by default; press the

SELECT key to change it to “

Capacity

Resistance    Diode 

   

Cont     Capacity 

(2) Connect the red test lead to V/

 terminal and the

 black test lead to COM terminal. 

(3) 

Connect the test leads to the both ends of the

capacitor under test. 

(4) Take the reading on the display. 

5-6 Temperature Measurement 

(1) Set the Function Switch to “

ºC/ºF

” position. 

(2) Connect the K-type Temperature Probe to the

input terminal. Positive (+) side of Probe should be

connected to V/

(3) 

Contact the Sensor (metal part) of K-type

Temperature Probe to the object under test. Take the

reading on the display. 

Positive (+) side of Probe should be connected to V/

WARNING 

 

Never connect the Temperature Probe to an

energized circuit. 

CAUTION 

 

Room temperature is indicated on the LCD when

 setting the Function Switch to “

ºC/ºF

” position. In

 

case that “

OL

” or anything other than room

 temperature is indicated, something may wrong

 with the instrument. Stop the use of instrument

 immediately. 

There may be a break in Probe when indication

 

isn’t changed if Sensor (metal part) of K-type

 Temperature Probe is contacted with the object

under test. 

6. Other functions 

6-1. Sleep Function 

(1) This is a function to prevent the instrument from being 

left powered on in order to conserve battery life. This 

function causes the instrument to enter Sleep mode 

about 15 minutes after the last key operation. To exit the 

Sleep mode, turn the Function switch to “

OFF

”, then to 

any other position. 

(2)  Sleep Function is disabled when; 
MIN/MAX or PEAK Function is selected. Continuous

measurement is made with the Sleep Function being

disabled. To activate Sleep Function again, disable the

MIN/MAX or PEAK Function. 

CAUTION 

 

The instrument consumes small amount of battery 

power in the Sleep mode. Set the Function Switch 

to the OFF position after use. 

6-2. HOLD Key

 

(1) Data Hold Function 

This is a function to freeze the measured value on the

display. Press the “HOLD” key to freeze the   reading. 
The reading will be held regardless of subsequent

variation in input. “

H

” is indicated on the upper left corner 

of the display while the instrument is in the Data   Hold

mode. To exit Data Hold mode, press the “HOLD” key

again. 

Anode

Cathode 

CAUTION 

 

Held readings are released when Sleep Function

is activated while the instrument is in the Data Hold 

mode. 

(2) Backlight ON/OFF 
Pressing the HOLD key 2 sec or more lights up the

Backlight. Pressing the HOLD key 2 sec or more again 

turns off the Backlight. 

6-3. NCV Function 

Red LED on the upper area on the Panel lights up 

at all ranges except for OFF when electric field   

exceeding 100V is detected by the sensor installed in 

the Jaws. It indicates a presence of voltage in an   

electrical circuit or equipment without 
touching them.NCV Sensor 
can detect electrical field   
only from the direction 
indicated in the right 
figure.Put the fixed   
element (left side) 
closer to the conductor 
under test. Detection against   
in-wall outlet is imposs ble.

 

DANGER 

 

The LED may not light up due to installation
 condition of electrical circuit or equipment. Never
 

touch the circuit under test to avoid poss ble

danger even if the LED for NCV doesn’t light up. 

 

Check the functionality of LED on a well-known
 power supply prior to measurement. When the
LED doesn’t light up, do not make measurement. 

 

NCV indication is affected by external voltage, how 

to hold or place the instrument. 

6-4. MIN/MAX Function 

 

CAUTION 

 

SELECT, ZERO, Hz/DUTY keys are disabled

while MIN/MAX Function is being activated. 

(1) 

AC/DC Current Range

 (600A only on DSA680) 

Pressing the MIN/MAX Key at 600A & 1000A Function

enables min or max value measurement. Press the

MIN/MAX Key to select MAX or MIN. The max or min

value within measuring range is being held until this

function is disabled. “MIN” or “MAX” is indicated on the 
display while this function is being activated. To disable

this function, press down the MIN/MAX Key at least 2

sec or change functions. 

(2) AC/DC Voltage Range 

CAUTION 

Pressing the MIN/MAX Key without applying voltage 

disables the Auto-ranging function and fixes the 

Range to 6V. Connect the test leads to the circuit 

under test and press the MIN/MAX Key after an 

appropriate range is selected by Auto-ranging 

function. 

Pressing the MIN/MAX Key enables min or max value 

measurement. Press the MIN/MAX Key to select MAX 

or MIN. The max or min value within measuring range is 

being held until this function is disabled. “MIN” or “MAX” 

is indicated on the display while this function is being 

activated. 
 To disable this function, press down the MIN/MAX Key 

at least 2 sec or change functions. 

6-5. ZERO Function 

 

CAUTION 

MIN/MAX, PEAK keys are disabled while ZERO 

Function is being activated. 

Zero Adjustment Function at Current Range “

” mark is 

to be indicated at the upper right on the display while 

ZERO function is being operated.   
Indication of relative value at Current, Voltage, 

Resistance: 
Pressing the ZERO Key indicates REL (relative value) 

Press the ZERO Key to save the initial value at the start 

of measurement as a reference value. Then the 

difference between the later measured values and the 

reference value is indicated on the display. The 

Auto-ranging function is disabled, while this function is 

being activated, and the Range is fixed to the Range 

selected at the start of measurement. Relative value is 

indicated within following ranges. 

(Measuring range) = 
(Full-scale value at the fixed Range) – (initial value) 

 To disable this function, press down the MIN/MAX Key 
 at least 2 sec or change functions. 

6-6. PEAK Function 

(600A only on DSA680) 

(1) Set the Function Switch to “

AC Current

” Range and 

 clamp onto a conductor under test. 
(2) Pressing the PEAK Key indicates “

P

MAX

” on the

 display and initiates measurement. 
(3) Readings indicates the PEAK of current crest value. 
 

When measuring sine wave, reading is about 

2

times of RMS value. 

(4) Press the PEAK Key at least 2 sec to reset the

 indication or release PEAK Function. 
Buzzer sounds twice, and the Function is released. 

CAUTION 

 

PEAK indication for Crest value is up to 1500A. 

Error indication is given when exceeding this

range value. 

 

Sleep Function is disabled when PEAK Function

is selected. Care should be taken when

performing 

continuous 

measurement.

6-7. Over-flow indication

 

When the input exceeds the measuring range at each 

Function other than Voltage ,1000A and Temperature 

Range "

OL

" or "

-OL

" is indicated on the display. 

7. Battery Replacement 

 

WARNING 

 

To avoid electrical hazard, set the Function Switch

 to  “

OFF

” and remove the test leads from the

 instrument before trying to replace batteries. 

CAUTION 

 

Do not mix old and new batteries. 

 

Install batteries in correct polarity as indicated in the 

Battery Compartment. 

Replace the batteries when a Low Battery Voltage warning 

BATT

” mark is indicated on the display. Note that when the 

battery is completely exhausted, the display blanks without 

BATT

” mark shown.

 

(1) Set the Function Switch to “OFF” position. 
(2) Unscrew and remove the Battery Compartment cover

on the bottom of the instrument. 
(3) Replace the batteries observing correct polarity. Use

new R03 (AAA) or LR03 / 1.5V batteries. 
(4) Install the Battery Compartment and tighten the   screws. 

8. Maintenance

 

 

Cleaning

Use a cloth dipped in water or neutral detergent for
cleaning the instrument. 
Do not use abrasives or solvents. Otherwise, instrument 
get damaged, deformed or discolored. 