A.W. Sperry Instruments 3005 Operating	 Instruction Download Page 10

Terminal connection of insulation resistance test

In case of testing insulation of insulated wire and cable against the earth
at direct current, connecting - pole of power to cable conductor, + to the
earth obtains smaller measuring value, compared with connecting the
other way round.  This connecting method is generally acknowledged
relevant to detect defective insulation.

6-3 Continuity Measurement (Resistance Tests)

DANGER

Always test the circuit or equipment to ensure it is surely de-energized
before measurement according to the instruction of 6-1.

To avoid electrical shock, measurements must be performed on de-

energized circuits only.

Never make measurement with the battery compartment cover

removed.

CAUTION

Never press the test button if the live circuit warning is indicated or the
warning buzzer sounds. This may damage the circuit.

When an additional circuit is connected in parallel with the circuit under
test , inaccurate reading may be taken.

Set the function switch to the "AUTO NULL" position.

Short the test probes (red) and (black) and press the test button. Then the

resistance of the test probes is displayed and memorized with
microprocessor.

Set the function switch to "

" position. 

Connect the test probes to the circuit under test and press the test button.

Read the resistance value from the LCD.

NULL symbol (

) is displayed while AUTO NULL function is working.

AUTO NULL will be cancelled when the instrument is powered off.

Principle of Continuity Measurement (Resistance Test)

Resistance value can be obtained by applying a certain current to the
resistance under test and measuring the voltage generated on the both
sides of the resistor under test.

Resistance value = Voltage / Current

RX        =   V    /   I

6-4 Continuous Measurement

A lock down feature is incorporated on the test button. Pressing and
turning it clockwise, lock the test button in the continuous operating
position.
To release the lock, turn the test button counterclockwise.

DANGER

Be extremely careful not to get electric shock during insulation resistance
measurement as high voltage is present on the tip of test probes
continuously.

17

16

Summary of Contents for 3005

Page 1: ...05 Form 270 OPERATING INSTRUCTION MODEL 3005 DIGITAL INSULATION CONTINUITY TESTER 99 Washington Street Melrose MA 02176 Phone 781 665 1400 Toll Free 1 800 517 8431 Visit us at www TestEquipmentDepot c...

Page 2: ...is essential to understand the safety rules contained in the manual They must be observed on using the instrument Be sure to observe the above rules strictly Not following the instructions may cause...

Page 3: ...take measurement in the presence of flammable gasses Otherwise the use of the instrument may cause sparkling which leads to an explosion Always keep your fingers behind the barrier on test probe durin...

Page 4: ...racy 1 5 rdg 5dgt 1 5 rdg 3dgt AC Voltage Indication 0 600V 5 rdg 3dgt 5 2 FEATURES MODEL 3005 are microprocessor controlled insulation continuity testers Designed to safety standards IEC 61010 1 Over...

Page 5: ...es 20 200 2000 280V DC AC p p for 10 seconds Range Protection by fusing 7 Typical Number of Measurements central tendency for supply voltage up to 8V Insulation Resistance Ranges Approx 1000 times min...

Page 6: ...ions 185 L x 167 W x 89 D mm approx Weight 900g approx including batteries Power Source 8 x R6P 1 5V AA or equivalent Auto power off Function Automatically turns off approx 10 minutes after the last s...

Page 7: ...S 4 3 Connector 11 4 2 LCD DISPLAY 1 INSULATION RESISTANCE SCALE 2 BAR GRAPH 3 CONTINUITY SCALE 4 LIVE CIRCUIT WARNING AC LIVE CIRCUIT DISCHARGE VOLTAGE WARNING 0 2V 0 2V 3 30V 3 60V 31 60V 61 120V 61...

Page 8: ...d to LINE terminal of connector terminal 12 6 OPERATION 6 1 Disconnection and check of power source of the circuit under test DANGER To avoid possible electrical shock do not perform measurements on e...

Page 9: ...l shock measurements must be performed on de energized circuits only When the test button is pressed with the range selector switch in the M position take care not to touch the tip of the test probe a...

Page 10: ...aken Set the function switch to the AUTO NULL position Short the test probes red and black and press the test button Then the resistance of the test probes is displayed and memorized with microprocess...

Page 11: ...ery compartment and replace the fuse Fuse type 600V 500mA F quick acting ceramic fuse 6 35 x 32mm 19 7 FUNCTIONS 7 1 AUTO NULL When conducting continuity tests the contact resistance of test probes et...

Page 12: ...act nature of the fault Before returning the instrument make sure that a Operating instructions have been followed b Leads have been inspected c Fuse has been checked d Battery has been checked e The...

Reviews: