Controls
Inputs / Observations
Adjust / Check
Observe voltage on pin 4
Check for -15 V
of the op-amp
Observe voltage on pin 7
Check for +15 V
of the op-amp
P
ROCEEDURE
B
:
Controls
Inputs / Observations
Adjust / Check
POWER: OFF
Remove op-amp
POWER: ON
Observe voltage on pin 4
Check for -15 V
of the op-amp socket
Observe voltage on pin 7
Check for +15 V
of the op-amp socket
POWER: OFF
Replace op-amp or
remove an additional
op-amp for further testing
POWER: ON
Zeroing
The following procedure calibrates Channel 1.
Controls
Inputs / Observations
Adjust / Check
POWER: ON
Connect
PROBE+
and
PROBE-
Adjust T100 for 0.0 V
to instrument
GND
Observe voltage at TP101
Observe voltage at TP102
Adjust T101 for 0.0 V
Observe voltage at TP103
Adjust T103 for 0.0 V
Repeat this proceedure for locations TP201 through TP203 and T200 through
T203 to calibrate Channel 2.