background image

- 2 -

- 3 -

Press the button once.

 

Twice

Press the button the specified number of 

times.

Slide the button to the 

right

.

Slide the button to the 

left

.

 

Hold

Hold the button in the direction of the arrow 

until the option you want appears on the 

screen.

 

Hold for

5 seconds

Hold the button in the direction of the arrow 

for the specified duration of time.

This manual uses the following graphic 

icons to show you how to operate the 

button.

Button Operations

2.6 Display

Delete flag

Control test flag

HYPO flag

After-exercise flag

Displays the 

blood symbol, 

date and time, 

etc.

Sound mute symbol
Before-meal flag

After-meal flag
Battery symbol

Clock counter

Meal amount flag

Thermometer symbol

2.7 Inserting Test Strips into the Meter

Start with the following steps to test your blood glucose level.

1. With clean, dry hands, remove 1 test strip from the bottle. 

Do not

 bend the test strip.

NOTE

n

 Tightly close the bottle immediately after taking out a test strip.

n

 

Do not

 apply blood to the test strip before inserting it into the meter.

2. Insert the test strip fully into the test strip slot. 

The meter turns on. Check that a full display appears as 

in chapter 2.6. 

After the most recent test result appears, the 

blood symbol will flash on the screen.

If nothing appears on the screen, remove the test strip and insert it back into 

the test strip slot.

NOTE

n

 You can turn on the test strip light.    

n

 The test result that appears right after power turns on is the most recent 

result without a control test flag or delete flag. 

n

 If the thermometer symbol appears on the screen, see “If a 

Message Appears” in the “Information for Users”.

n

 If you insert a test strip beyond the expiration date you set, the 

blood symbol and expiration symbol will alternately flash on the 

screen. If this occurs, see chapter 6.3.

CAUTION

If you cannot test your blood glucose due to a problem with your meter or 

test strips, promptly contact your healthcare professional and distributor.

3. To flag the test result: 

See step 2 in chapter 5.1.                     

Next ... 

Go to chapter 3 “Blood Sampling”. Draw blood and apply it to the test strip no 

more than 5 minutes after you insert the test strip into the meter. Double beeps 

begin to sound 4 minutes and 30 seconds after inserting the test strip. Thirty 

seconds later, the meter turns itself off.

NOTE

In the ECO mode, the blood symbol flashes 2 minutes shorter than otherwise. 

See chapter 7 “Meter Setup”.

Chapter 3  Blood Sampling

This chapter briefly explains how to get a blood sa

mple using the Glucoject

®

 

Dual / Dual S

 

and Glucoject

®

 Lancets No-dol. If using some

 other lancing 

device to sample blood, see the manual for that device. 

3.1 Lancing Device Parts

3.2 Cautions on Using the Glucoject

®

 Dual / Dual S

Follow these cautions to reduce the risk of infection by pathogenic 

microorganisms.

 

CAUTION

n

 Wash the puncture site with soap and water. Dry the site thoroughly 

before sampling blood. 

n

 

Do not

 share the same lancet or lancing device with anyone to avoid the 

risk of infection. 

n

 Always use a new lancet. Lancets are for single-use only. 

Do not

 reuse a 

lancet that you have already used.

3.3 Selecting a Sampling Site

Alternative site testing (AST)

: This meter can test the glucose level of blood 

from your fingertip, palm, forearm or upper arm. However, test results from 

sites other than the fingertip may differ from a fingertip measurement. Consult 

your doctor or healthcare professional before testing blood from the palm, 

forearm or upper arm.

Use blood from:

If you are going to test:

Fingertip, palm, 

forearm or upper 

arm

n

 Before meals

n

 Two hours or more after meals

n

 Two hours or more after exercise

Fingertip

n

 When there is the possibility of blood glucose level 

changing rapidly (e.g. after meals or exercise) 

n

 When experiencing symptoms of hypoglycemia 

such as perspiration, cold sweats, a floating 

sensation or trembling 

n

 When immediate testing is needed for suspected 

hypoglycemia 

n

 When in poor physical condition such as having a 

cold, etc.

3.4 Sampling from Your Fingertip

1.

  Remove the head cap. 

2

.  Insert a new lancet into the holder until it stops. 

NOTE 

Use a lancet before its expiration date. 

3

. Twist off the protective cap on the lancet. Keep the 

protective cap for disposing of the lancet after the test. 

Glucoject

®

 Lancets No-dol

Glucoject

®

 Dual / Dual S

Protective cap

Depth adjuster

Penetration depth indicator

Green release 

button

Device sleeve

Clear alternative site 

end cap

Grey force 

adjustment feature

Left                  Right

Depht adjuster collar

Summary of Contents for GLUCOCARD MX

Page 1: ...induce severe low glucose symptoms such as coma 2 2 What You Need for Each Test a GLUCOCARDTM MX meter b GLUCOCARDTM MX BLOOD GLUCOSE TEST STRIPS c Lancing device d Lancet WARNING Keep the meter test...

Page 2: ...test strip into the meter Double beeps begin to sound 4 minutes and 30 seconds after inserting the test strip Thirty seconds later the meter turns itself off NOTE In the ECO mode the blood symbol flas...

Page 3: ...d The meter counts down from 5 to 1 during the test NOTE n For accurate test results touch the test strip to your blood in less than 20 seconds after puncturing n Do not test blood that runs or spread...

Page 4: ...yond the 500th overwrite previously stored results in the order of oldest first 1 Make sure the meter is off 2 Turn on the meter The most recent test result appears on the screen The 3 digit number fo...

Page 5: ...ucose When the clock counter setting is on the elapsed time and after meal flag appear on the right of the test result 2 To change or set the elapsed time Select the time Confirm your setting 3 Slide...

Page 6: ...old To make another alarm setting repeat steps 1 to 4 To the top of the beeper volume setting To the top of the date setting The meter turns off NOTE The meter turns itself off if left alone for 3 min...

Page 7: ...toring Your Meter After use tightly close the caps of the test strip bottle and control solution bottle to maintain their quality Store your meter test strips control solutions and manuals in your car...

Page 8: ...accuracy Within 75 seconds month Operating Temperature 10 to 40 C 50 to 104 F environment Humidity 20 to 80 RH No condensation Expected life 3 years according to company data Dimensions L102 W33 18 5...

Page 9: ...Operating Manual MASTER 43125 12 11 420 x 300 mm BLACK K 100...