User manual ABE_300
Biogas Chiller ABE_300
23
4.7
How to calibrate analog input
For the calibration of each channel to proceed with the following steps:
1. Connect the analogue input signal to zero take on zero value of the channel to write the value of zero.
2. Press the left arrow to capture and combine the reading with the value introduced by the converter.
3. Connect the analogue input signal span bring the value of the channel span to write the value of span.
4. Press the left arrow to capture and bind the value read from the converter with the given span of
introduced.
5. Press the fan to store the changes in non-volatile memory and confirm by pressing 2 times the enter key.