20
Guidance and manufacturer’s declaration -electromagnetic emissions
The LED Curing Light LED-3200 is intended for use in the electromagnetic environment
specified below. The customer or the user of the LED Curing Light LED-3200 should
ensure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment-guidance
RF emissions
CI SPR I 1
Group 1
The LED Curing Light LED-3200 uses
RF energy only for its internal function.
Therefore, its RF emissions are very low
and are not likely to cause any interfe-
rence in nearby electronic equipment.
RF emissions
CISPR11
Class B
The LED Curing Light LED-3200 is
suitable for use in all establishments,
including domestic establishments and
those directly connected to the public
low-voltage power supply network that
supplies buildings used for domestic
purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies
Recommended separation distances between portable and mobile RF communications
equipment and the Medical LED Curing Light LED-3200
The LED Curing Light LED-3200 is intended for use in an electromagnetic environment
in which radiated RF disturbances are controlled. The customer or the user of the Medical
LED Curing Light LED-3200 can help prevent electromagnetic interference by maintai-
ning a minimum distance between portable and mobile RF communications equipment
(transmitters) and the LED Curing Light LED-3200 as recommended below, according to
the maximum ouput power of the communications
Rated maximum
outputpower of
transmitter w
Separation distance according to frequency of transmitter m
150 KHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2,5 GHz
0.01
0. 1 2
0.1 2
0.23
0. 1
0.38
0.38
O. 73
1
1.2
1.2
2.3
1 0
3.8
3.8
7.3
1 00
1 2
1 2
23
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in meters(m) can be estimated using the equation applicable to
the frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watts(W) according to the transmitter manufacturer.
NOTE 1 At 80MHz and 800 MHz, the separation distance for the higher frequency
range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.