Pressure Infusor
Refer to the Operator’s Manual for
warnings, instructions and other
information related to this product.
7.
Press the Start/Stop button to turn the
corresponding infusor chamber ON. The
“in Range” LeD will flash green while the
system is pressurizing.
8
. When the “in Range” LeD turns solid green,
open clamps to begin flow.
9
. When fluid bag is depleted, press the
Start/Stop button to turn the pressure
infusor chamber OFF.
Troubleshooting
Standby/ON mode
Start/Stop buttons
Low
High
Low
High
In Range
In Range
300 mmHg
3M
™
Ranger
™
Pressure Infusor
Condition
Cause
Solution
Nothing illuminates on
the pressure infusor
control panel when the
main power switch is
turned ON.
Power cord is not plugged into the
power entry module of the unit, or
power cord is not plugged into a
properly grounded outlet.
Make sure the power cord is plugged into the
power entry module of the pressure infusor.
Make sure the pressure infusor is plugged
into a properly grounded outlet.
Unit failure.
Contact a biomedical technician.
Power LED status
lights do not
illuminate.
Power cord is not plugged into the
power entry module of the unit, or
power cord is not plugged into a
properly grounded outlet.
Make sure the power cord is plugged into the
power entry module of the pressure infusor.
Make sure the warming unit is plugged into a
properly grounded outlet.
Unit is not turned ON.
Using the main power switch located under
the pressure infusor, turn the unit ON.
Burned out LeD light.
Push the pressure infusor power button.
if unit functions properly, continue use.
Contact a biomedical technician after use to
replace LeD light.
Unit failure.
Contact a biomedical technician.