3M Ranger Quick Start Manual Download Page 2

3M

 Ranger

 Pressure Infusor 

Refer to the Operator’s Manual for 

warnings, instructions and other 

information related to this product. 

7. 

Press the Start/Stop button to turn the 

corresponding infusor chamber ON. The 
“in Range” LeD will flash green while the 
system is pressurizing. 

8

. When the “in Range” LeD turns solid green, 

open clamps to begin flow. 

9

. When fluid bag is depleted, press the 

Start/Stop button to turn the pressure 
infusor chamber OFF. 

Troubleshooting 

Standby/ON mode 

Start/Stop buttons 

Low

High

Low

High

In Range

In Range

300 mmHg 

3M

 Ranger

™ 

Pressure Infusor 

 Condition

 Cause

 Solution 

Nothing illuminates on 

the pressure infusor 

control panel when the 

main power switch is 

turned ON. 

Power cord is not plugged into the 

power entry module of the unit, or 

power cord is not plugged into a 

properly grounded outlet. 

Make sure the power cord is plugged into the 

power entry module of the pressure infusor. 

Make sure the pressure infusor is plugged 

into a properly grounded outlet. 

Unit failure. 

Contact a biomedical technician. 

Power LED status 

lights do not 

illuminate. 

Power cord is not plugged into the 

power entry module of the unit, or 

power cord is not plugged into a 

properly grounded outlet. 

Make sure the power cord is plugged into the 

power entry module of the pressure infusor. 

Make sure the warming unit is plugged into a 

properly grounded outlet. 

Unit is not turned ON. 

Using the main power switch located under 

the pressure infusor, turn the unit ON. 

Burned out LeD light. 

Push the pressure infusor power button. 

if unit functions properly, continue use. 

Contact a biomedical technician after use to 

replace LeD light. 

Unit failure. 

Contact a biomedical technician. 