MAINTENANCE
087-0065, Rev D/February 2019
6-10
6. Reinstall the sensor.
7. Install the sensor cover and tighten the screws with the sequence in
.
)
6.2.3. REPLACING THE SENSOR FILTER
Replace sensor filters as follows:
1. Refer to steps 1 through 3 in
Section 6.2.1. REPLACING SENSORS
to remove the sensor cover.
2. Pull out the old sensor filter with tweezers or other tools as shown in
Figure 6-9: Removing the Old Filter
3. Discard the old sensor cover.
4. Replace with the new sensor filter (see
).
5. Reinstall the sensor cover and tighten the screws with the sequence illustrated in
WARNING: AFTER REPLACEMENT, PERFORM A BUMP TEST AND CALIBRATION TO
ENSURE PROPER OPERATION PRIOR TO PLACING BACK INTO SERVICE.
FAILURE TO FOLLOW THIS WARNING COULD LEAD TO SERIOUS
INJURY OR DEATH.
WARNING: DO NOT REPLACE FILTERS WHILE IN A HAZARDOUS LOCATION.
FAILURE TO FOLLOWING THIS WARNING COULD LEAD TO SERIOUS
INJURY OR DEATH.
CAUTION: Do not attempt to replace the individual filters. Doing so violates ingress
protection rating and product certification.
CAUTION: 3M recommends wearing rubber gloves when handling the new sensor filter
to avoid damage from fingernails or natural skin oils.
WARNING: AFTER REPLACEMENT, PERFORM A BUMP TEST AND CALIBRATION TO
ENSURE PROPER OPERATION PRIOR TO PLACING BACK INTO SERVICE.
FAILURE TO FOLLOW THIS WARNING COULD LEAD TO SERIOUS
INJURY OR DEATH.
Sensor Cover
Filter