Storage and maintenance
•
Disconnect the warming unit from the power source before cleaning.
•
Wipe the outside of the warming unit with a damp, soft cloth and a mild detergent or a hospital approved
hard surface disinfectant.
•
Wipe with a dry cloth.
•
Cleaning the labels with alcohol will over time cause the labels to fade.
•
Do not use solvents. These will damage the case and labels.
•
The Ranger warming plates can be periodically cleaned using the 3M
™
Ranger
™
Cleaning Tool.
•
Unplug the warming unit from its power source.
•
Unfold the cleaning tool and wet the foam pads with non-abrasive detergent or cleaning solution.
•
Insert the tool from the back of the unit until it protrudes from the front of the unit.
•
Pull the cleaning tool all the way through from the front of the unit.
•
Rinse the tool with water and repeat, passing it through the warming unit 3 times.
•
Wipe any excess fluid from the warming unit.
The 3M
™
Ranger
™
Blood and Fluid Warming Sets
•
Six styles of warming set are available in the UK:
One neonatal/paediatric (KVO-6L/hr), two standard flow (KVO-9L/hr) and three high flow
(KVO-30L/hr) disposable sets to meet your fluid warming needs.
•
All warming sets incorporate an air capture system, to reduce the risk of an air emboli being infused into the patient.
•
All high flow sets are equipped with an auto-venting bubble trap, that is capable of venting up to 3 litres
of air per minute.
•
Sets can only be inserted into the warming unit one way, this prevents incorrect placement.
•
All sets can be pressurised to 300 mmHg.
•
All sets are packed sterile and are latex and DEHP free.
Stand alone pressure infuser
•
When mounting the pressure infuser to an IV
pole, position it no higher than 142cm (56”) off
the ground, on to a heavy based five footed
IV pole. This will reduce the risk of the IV pole
tipping and potentially causing injury
to someone.
•
Secure the infuser to the IV pole by tightening
the two clamps on the back of the unit and the
power cable, using the hook-and-loop strap.
•
Never infuse fluids if air bubbles are present
in the fluid line, air embolism may result.
•
The pressure infuser can be used with all
3M
™
Ranger
™
Blood and Fluid Warming Sets,
as well as 3M
™
Ranger
™
Irrigation Sets and will
accept fluid bags from 250mL to 1000mL
in volume.
Continued on the next page