Read Before Servicing Equipment
The repair, calibration, and servicing of the warming unit
requires the skill of a qualified medical equipment service
technician who is familiar with good practice for medical
device repair. If service is designated as not requiring
manufacturer’s attention, the technical information is provided
in this service manual or will be provided, on request, by 3M
Patient Warming
Refer to Service Manual
Perform all repairs and maintenance in accordance with the
instructions in this service manual.
Safety Inspection
Perform a safety inspection after making repairs to the
Bair Hugger warming unit and before returning the warming
unit to service. A safety inspection should include a test of the
operating temperatures (described in this service manual), the
Over Heat alarm system, as well as a leakage current test.
Proper Use and Maintenance
3M Patient Warming assumes no responsibility for the
reliability, performance, or safety of the equipment if:
• Modifications or repairs are performed by non-authorized
personnel.
• The equipment is used in a manner other than that described
in the Operator’s or Service Manuals.
• The equipment is installed in an environment that does not
meet the relevant grounding requirements.