NOTICE
PURCHASER’S RESPONSIBILITY:

Only health care professionals or other appropriately trained individuals should use this equipment.

Parts Identification

1. Disposable Blade Assembly (9670)

2. ON/OFF Switch

3. Clipper Body (9671)

4. Indicator Light

5. Charger

The 3M Surgical Clipper with Fixed Head 9671 must be used with one 
of the following 3M Chargers:

3M Model

Voltage (AC)

9672

Charger

120V, 60Hz US/Canada

9673

Charger

240V, 50Hz Australia/New Zealand (Not UL Approved)

9674

Charger

100V, 50-60Hz Japan (Not UL Approved)

9675

Charger

230V, 50Hz United Kingdom/Ireland

9678

Charger

230V, 50Hz Europe

Clipper Specifications

Transformer Output:

1.3V, 1.91A

Motor Voltage:

1.2 Volts DC

Current without load: 0.58 amps max.

Current with load: 3.18 amps max.

RPM: 5800-8600

Charge Current:

160-210 milliamps

Running Current:

1.2-1.5 amps

Charging Time:

18-24 hours initial, 12-14 hours thereafter

Battery Operating Time:  

~45 minutes

Clipper Weight:

201 grams (7.1 oz)

Clipper Dimensions:

14.1 cm (5.55 in) Height

3.8 cm (1.5 in) Width (blade) 

3.7 cm (1.46 in) Width (body) 

3.4 cm (1.34 in) Depth 11.3 cm (4.45 in) 

10.8 cm (4.25 in) Circumference

INSTRUCTIONS FOR USE

To Attach a Disposable Blade Assembly onto the Clipper Body:

Make sure that the ON/OFF switch is in the OFF position. Hold the clipper body at a 45° (degree) angle over the opened blade assembly package. Push the clipper
body into the blade to secure. An audible snap indicates blade assembly is securely attached (A). 

To Use the Clipper:

The patient’s skin should be clean. To turn the clipper on, slide the ON/OFF switch to the ON position. Use short, gentle strokes, holding the clipper blade assembly at
a 15 to 30 degree angle to the patient’s skin (B).  For best results, clip against the direction of the hair growth. For problem areas, stretch the skin taut while clipping. 

To Remove Disposable Blade Assembly:

Assure the ON/OFF switch is in the OFF position. Position the clipper over an appropriate waste receptacle with the blade assembly facing downward. Use your
thumb to push blade assembly off the clipper body into waste receptacle (C).

Care and Cleaning of the Clipper Body:

The housing of the clipper body is made of durable plastic and may be wiped off with a damp cloth containing a small amount of alcohol, disinfectant, or a mild
detergent solution. While uncorded, the clipper body may also be rinsed under running water (D). Allow to dry completely before reconnecting cords or reusing. 
Do not immerse clipper body in any liquid (E). 

To Charge the Clipper:

1. Slide the ON/OFF switch to the OFF position.

2. Clean the clipper body before reconnecting to the cord, making sure the clipper is free of hair or other contaminants.

3. Connect the cord to the receptacle in the bottom of the clipper and then to an appropriate electrical outlet (120 volts AC 50-60 Hz), making sure the cord is not

pinched or kinked or damaged. 

4. The charging indicator light will be on when the clipper is plugged in and making the proper connection. If the charging indicator light fails to illuminate when the

clipper body is plugged in, make sure the cord has proper contact with the clipper terminals and it is free of hair or debris.

5. To be certain that the clipper is fully charged, plug the clipper in for 18 to 24 hours before its initial use. Thereafter, the clipper can be fully charged in 

12 to 14 hours.

6. We recommend leaving the clipper plugged in between uses. It cannot be overcharged. 

A

B

C

D

E

Summary of Contents for 9671

Page 1: ...ipper med fast hoved Kirurgisk clipper Kkirurgiset karvanleikkurit kiinteällä päällä Máquina eléctrica de tricotomia com cabeça fixa ÃÂÈÚÔ ÚÁÈÎ ÚÈÛÙÈÎ ªË Ó Ì ٠ıÂÚ ÂÊ Ï 9671 3 Clipper Made in USA from Foreign and US Components for 3M Health Care St Paul MN 55144 1000 Charger Made in China for 3M Health Care St Paul MN 55144 1000 Blade Assembly Made in USA for 3M Health Care St Paul MN 55144 1000 ...

Page 2: ...icient power to remove course hair yet is gentle enough for fine hair It works well on all types of hair either wet or dry and in delicate areas For best performance clip hair dry The clipper has been designed and tested for use only with the 3M Surgical Clipper with Fixed Head model 9671 series battery chargers 9672 9673 9674 9675 9678 and the 3M disposable blade assembly 9670 WARNING To reduce t...

Page 3: ...pper blade assembly at a 15 to 30 degree angle to the patient s skin B For best results clip against the direction of the hair growth For problem areas stretch the skin taut while clipping To Remove Disposable Blade Assembly Assure the ON OFF switch is in the OFF position Position the clipper over an appropriate waste receptacle with the blade assembly facing downward Use your thumb to push blade ...

Page 4: ...AM to 6 00 PM Central Time After business hours a voice prompt system will take your product complaint information For customers outside of the United States please contact your local 3M Sales Representative for necessary shipping labels and packing list Please include a description of the defect with the returned unit 3M Health Care Attn Complaint Coordinator 3M Center Bldg 270 5N 01 St Paul MN 5...

Reviews: