2
2. Product Description
The RotoMix unit is a mixing device for dental filling
materials and cements in capsules. Unlike conventional
capsule mixers, the device uses a rotating motion which
is faster and more efficient and leads to improved
homogeneity of the mixed pastes. In addition, RotoMix
offers the option to centrifuge the capsules after mixing.
Depending on the viscosity this procedure may reduce air
bubbles in the material. Desired mixing and centrifugation
times may be programmed into the mixing programmes
“P1”-“P4”. Moreover, two pre-set mixing times, “10 s” and
“8 s”, can be selected.
Operating Instructions should not be discarded for the
duration of product use.
3. Technical Data
Drive unit:
Ball-bearing motor with
integrated self-resetting
overheat cut-off
Operating mode:
Short-term operation, 2 min
On/5 min Pause (to prevent
needless motor heating)
Controller: Microprocessor-controlled,
high-precision mixing time
control by quartz timing
mechanism
Capsule holder:
Self-locking
Capsules:
Diameter max. 19 mm,
total length max. 58 mm,
length without tubes
max. 38 mm,
weight max. 6 g
Housing:
Housing and protective cover
consist of high impact-resistant
plastics
Mixing speed:
typically 2850 for 6 g-capsules
Centrifugation speed:
typically 2950 for 6 g-capsules
Operating temperature: 18°C…40°C/64°F…104°F
Humidity:
max. 80% at 31°C/88°F,
max. 50% at 40°C/104°F
Main voltage:
120 V, 230 V, 50/60 Hz,
for voltage setting refer to the
rating plate
Power requirements:
390 VA
Fuses:
230 V: 2 A time-lag fuse
120 V: 3.2 A time-lag fuse
Dimensions:
(W x H x D)
180 x 210 x 250 mm
Weight:
4000 g
Continous sound
pressure level:
= 73.8 dB(A)
Year of manufacture:
See rating plate
4. Installation
Prior to the first start-up remove the transport safeguard
(foam ring) and instruction card.
Connect the RotoMix unit to the power supply and switch
on the unit with the power switch on the back of the devi-
ce. The display shows the mixing time of programme 1.
The following basic settings are factory-set:
Programme 1 (key “P1”):
12 sec mixing,
3 sec centrifugation
Programme 2 (key “P2”):
9 sec mixing,
3 sec centrifugation
RotoMix
TM
Protective cover
Capsule
holder
Revolving
platform
Power
switch
Control panel
왗
14. Do not expose the device to solvents, flammable
liquids, and strong heat sources as these may damage
the plastic housing.
15. Cleaning the unit, ensure that no cleaning agent
enters the device since this may lead to electric short
circuiting or hazardous malfunctions.
16. Pay attention to capsule leak tightness when mixing.
Capsules that are not leak tight contaminate the device
with mercury. Contaminated devices cannot be
repaired and must be disposed of as contaminated
waste according to local laws.
See the safety data sheet for the amalgam used for
first aid, measures to take in case of unintended
release, handling, storage, exposure limitation, and
personal protective clothing.