59
Laser
Laser On:
Toggles the laser on or off, by either check (on) or uncheck (off).
Shutter:
Toggles safety shutter open or closed, by either check (open) or uncheck (closed).
Small Spot:
Sets IPM to the small spot focus position.
Large Spot:
Sets IPM to the large spot focus position.
Set Laser Power:
Sets Laser Power (in milliwatts mW) to the value that is in the power field, if
achievable.
Get Power at Vat:
Directs beam to the Full Area Detector (FAD) and determines the power, in
milliwatts, as measured by the FAD.
Recalculate Power Coeff:
Adjusts the power calibration for the decline of laser power over time.
This option is available here, but is not necessary for the user to perform. This function is
accomplished each time the laser power is set.
70696 Revision 1C 27-Aug-08