Pannoramic DESK 1.18.1 User’s Guide
5 Troubleshooting and Maintenance
Slide is often positioned incorrectly
•
Guiding fault in transport system
•
Contact 3DHISTECH Service for support.
Digitization process is extremely slow
•
Extended focusing
option is activated
•
Deactivate the
Extended focusing
option if not required.
•
Z-stack
option is activated
•
Deactivate the
Z-stack
option if not required.
Image is out of focus
•
Focus range limit is activated
•
Deactivate the
Limiting focus range
option.
•
Focus range setting is incorrect
•
Change focus range setting.
All individual images show a shading effect
•
Algorithm failed to find empty image fields
•
Capture a reference image with a blank slide (see section
).