3D Histech Pannoramic DESK 1.20 User Manual Download Page 47

Pannoramic DESK 1.20 User’s Guide

4 Software Description

Navigation Tool:

This tool offers easy and punctual slide navigation in distictive steps. Set 

Step Size

 (FoV) – 

1

½

, or 

¼

 to determine the size of movement in proportion of one FOV. Click an arrow to 

move to the required direction with the set step size.

Marker Radius

 – Set 

3

2

, or 

1

 as the radius of the marker, then mark area to be scanned by 

clicking the + button.

You can select the area to be scanned with the different tools on the 

Preview toolbar

:

 

Square tool

 

Ellipse tool

 

Polygon tool

: you can finish the selection process with the polygon tool by 

double-clicking with the mouse.

 

TMA tool

To undo your last action, click 

 (maximum of 40 changes).

To position the specimen holder that holds the current slide, click 

Position sample

 

 

and 

then click on the area of interest on the preview image. This will define the Live image that 
you can access from the Focus panel.

To display the original image created by the preview camera, click 

.

To shrink or enlarge the preview image, set the zoom level at the left 

Size

 selection box 

(100%-600%). To fit the preview image to the preview area, click 

Fit

.

Choose from the 

Scanned

 

 

or the 

Not scanned

 

 option from the right selection box. 

To add the selected areas to the preview, select the 

Scanned

 

 

option. 

To delete the selected areas from the preview, select the 

Not scanned

 

 option.

Figure 27 – Not scanned selection

October 9, 2015 - Rev. 1

3DHISTECH Ltd.

47

(

76

)

Summary of Contents for Pannoramic DESK 1.20

Page 1: ...PD120 UG Rev1 Pannoramic DESK 1 20 User s Guide October 9 2015 Rev 1 ...

Page 2: ...ns 16 1 2 3 Control Software 17 2 Installation 19 2 1 Control computer ports 19 2 2 Connectors and cables 20 3 Preparing Slides 21 3 1 Affixing Barcode Stickers to Slides 22 3 2 Inserting and Removing Slides 23 4 Software Description 26 4 1 Starting Exiting Pannoramic DESK Software Application 26 4 1 1 Starting the software 26 4 1 2 Exiting the software 26 4 2 Options 26 4 2 1 Barcode Settings 27 ...

Page 3: ... Server 56 4 5 Using Profiles 61 5 Troubleshooting and Maintenance 62 5 1 Safety Information 62 5 1 1 Cleaning the device 62 5 1 2 Preventive maintenance 63 5 2 Troubleshooting 63 5 3 Maintenance 66 5 3 1 Removing the protective cover 66 5 3 2 Replacing the halogen lamp 67 5 3 3 Removing a slide 69 5 3 4 Transporting Pannoramic DESK 70 6 Technical Data 72 Index 76 October 9 2015 Rev 1 3DHISTECH Lt...

Page 4: ...P are the sole properties of the 3DHISTECH Ltd 3DHISTECH Ltd is not liable for damage of whatever nature including but not limited to general or specific damage indirect damage consequential damage or incidental damage including the results of the analysis of the digitized slides for example change of health status related to erroneous diagnosis from the digitized slide s that stems from or is ass...

Page 5: ...IND EXPRESSED IMPLIED OR OTHERWISE INCLUDING WITHOUT LIMITATION ANY WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE IN NO EVENT SHALL SAM LEFFLER OR SILICON GRAPHICS BE LIABLE FOR ANY SPECIAL INCIDENTAL INDIRECT OR CONSEQUENTIAL DAMAGES OF ANY KIND OR FOR ANY DAMAGES WHATSOEVER RESULTING FROM LOSS OF USE DATA OR PROFITS WHETHER OR NOT ADVISED OF THE POSSIBILITY OF DAMAGE AND ON ANY...

Page 6: ...1 Particular requirements for in vitro diagnostic IVD medical equipment EN 55011 1998 Group 1 Class A Limits and methods of measurement of radio disturbance characteristics of industrial scientific and medical ISM radio frequency equipment This product has been tested to the requirements of CAN CSA C22 2 No 61010 1 second edition including Amendment 1 or a later version of the same standard incorp...

Page 7: ...h as F11 or the Ctrl M key combination Example Cross references within this document or to other documents Warning It might cause irreversible damage or harm to the user of the product or the product itself if this instruction is not followed Caution It might cause the product to not work properly if this instruction is not followed and alerts to a potential danger situation Note Calls attention t...

Page 8: ... All claims for warranty and warranty services will be null and void in any such case Warning Insert the power cord plugs only into a grounding outlet The protective action must not be rendered ineffective by extension cables without protective conductor The Pannoramic DESK has a built in protection which must not be rendered ineffective by extension cables without protective earth ground conducto...

Page 9: ... on Caution Avoid touching the lamp bulb Disconnect power plug and let the lamp cool down for about 15 minutes before replacing the lamp Crush hazard Never open the front door unless prompted accordingly by software When you open the door at any other time make sure not to reach with your hands into the instrument to avoid the risk of crushing your fingers Warning The front door is equipped with a...

Page 10: ...cepted in cases of natural wear notably wear and tear parts or improper product handling The Product Manufacturer will refuse any liability in the event of damage resulting from faulty operation negligence or unauthorized intervention into the product notably from the replacement of product parts or the usage of accessory parts from other manufacturers Any such case will void the user s claim for ...

Page 11: ...rtual Slide Slide A digital image of a thin glass plate on which specimens are mounted for microscopic study a dynamic interactive image that you can manage save magnify zoom name evaluate annotate mark and comment send to a colleague electronically for co operation or advice and so on October 9 2015 Rev 1 3DHISTECH Ltd 11 76 ...

Page 12: ...ill highly affect the functionality of the scanner program 1 1 2 Features and Benefits The Pannoramic DESK offers world class brightfield whole slide scanning An outstanding 0 12 µm pixel resolution is achieved with the 40 0 95 NA equivalent to 80 magnification Carl Zeiss Plan Apochromat objective Key features of the Pannoramic DESK Scanner are 3MP camera for barcode reading and preview image Impr...

Page 13: ...scription 1 1 3 Warning and Information Labels The following figure contains the labels used on the Pannoramic DESK main unit and the external mains power supply unit Figure 1 Warning and information labels October 9 2015 Rev 1 3DHISTECH Ltd 13 76 ...

Page 14: ...Product Description 1 1 4 Base Unit The following figure shows the main parts of the base unit Figure 2 Main parts of the base unit 1 Protective cover 2 Protection rail 3 Scan camera 4 Sliding door October 9 2015 Rev 1 3DHISTECH Ltd 14 76 ...

Page 15: ... following hardware and software items 1 2 1 Pannoramic DESK system The following figure shows the main components of the Pannoramic DESK system Figure 3 Pannoramic DESK system overview 1 24 V external power supply unit 2 Pannoramic DESK base unit 3 Monitor 4 Mouse 5 Control computer 6 Keyboard October 9 2015 Rev 1 3DHISTECH Ltd 15 76 ...

Page 16: ... monitor can vary depending on the product order Cameras Stingray F146C IRF Medical camera 1 CCD camera for normal brightfield application CIS VCC FC60FR19CL CMOS camera for high quality brightfield applications allowing fast scanning Camera adapters 0 63 adapter 1 adapter 1 6 adapter October 9 2015 Rev 1 3DHISTECH Ltd 16 76 ...

Page 17: ...rson Base Unit options Pannoramic DESK Basic Scan Option Fast Option base unit equipped with Stingray F146C IRF or CIS VCC FC60FR19CL camera for Brightfield scanning Objectives Carl Zeiss Plan Apochromat 20 NA 0 8 Carl Zeiss Plan Apochromat 40 Corr NA 0 95 Warning Do not change correction collar settings on 40 NA objective if you need it to be set contact customer service 1 2 3 Control Software Pa...

Page 18: ...ramic Viewer software fully functional Optional CaseCenter Pannoramic Viewer TMA module 3D Reconstruction module QuantCenter HistoQuant NuclearQuant MembraneQuant and DensitoQuant modules TumorBoard module E School Encoder October 9 2015 Rev 1 3DHISTECH Ltd 18 76 ...

Page 19: ...installation carried out by an unauthorized person Only qualified service assistants are allowed to install the hardware and software The product is ready for use after installation 2 1 Control computer ports The following figure shows the location of the connectors in order to enable easier transportation of the product for a short distance by disconnecting cables and detaching hardware component...

Page 20: ...mputer Preview camera USB cable between scanner and control computer Stingray FireWire 1394 B cable between Stingray camera and control computer CIS a pair of CameraLink cables between CIS camera and control computer Video cable between monitor and control computer Power supply cable between 230V 110V and monitor Power supply cable between 230V 110V and control computer Power supply cable between ...

Page 21: ... when it is broken or its edges are not properly cut or grounded Using such slides can lead to injury Warning Ensure that no part of the mounted coverslip protrudes from the edges of the slide Use slides with completely dried embedding medium only Otherwise the coverslip can be shifted that makes slide insertion less reliable Not completely dried embedding might also influence the operation of obj...

Page 22: ...code sticker 2 from the roll 2 Stick it on the colored label area 1 of the slide on the specimen side Caution Ensure that a marginal space of 1 2 mm remains on all sides between the sticker and the label area outer limits It is vital for barcode identification Do not let barcode sticker protrude from slide edges or stick on the coverslip surface 3 The thickness of the barcode sticker on the slide ...

Page 23: ...nd its removal can lead to injury and contagion Broken slide can be removed only by a well trained person Figure 6 Inserting and removing a slide 1 Click in Pannoramic DESK control software The specimen holder moves to the load unload position 2 Open the slide insert door 1 by pushing it to the left Caution When opening the slide insert door fingers may be pinched during operation 3 Open the fixin...

Page 24: ...ion Broken slide can be removed only by a well trained person 5 Close the fixing lever 2 by pushing it to the right so that it touches and therefore fixes the slide 6 Close the slide insert door 1 by pushing it to the right Caution When closing the slide insert door fingers may be pinched during operation Warning Never open the front door unless prompted accordingly by software When you open the d...

Page 25: ... slide insert door 1 by pushing it to the left 3 Open the fixing lever 2 by pulling it to the left 4 Remove the slide 4 5 Insert another slide or close the fixing lever 2 and the door 1 Note The slide insert door is equipped with an interlock switch It automatically stops a running digitization process when the slide insert door is opened October 9 2015 Rev 1 3DHISTECH Ltd 25 76 ...

Page 26: ...uccessful initialization the program displays a Mode selection window Figure 7 Mode selection window The selection window displays Manual Brightfield scan mode for digitizing individual slides If you had exited this mode and want to reactivate it click Manual 4 1 2 Exiting the software If you close a mode or one of the options window the Mode selection window turns active again To exit the softwar...

Page 27: ...ning the barcode type foreground color black or white the color of the barcode and other settings you enable the Read Barcode option when using the control software Note You can only scan the barcode type that you have previously set If you want to read several barcode types you always have to set the current barcode type before reading it October 9 2015 Rev 1 3DHISTECH Ltd 27 76 ...

Page 28: ...e the barcode reading area by dragging the edges of the green square To check whether the barcode reading settings are correct display the barcode digits in the Barcode field by clicking Note If you did not set the barcode type correctly or there is no barcode in the barcode area the software will display the following error message Barcode detection unsuccessful There might be no barcode present ...

Page 29: ...ou select Exclude check digit then the check digit number will also be analyzed as a barcode number Foreground Color You can select the following foreground color of the barcode area Black White Barcode Image Compensation Click Use barcode compensation image to use the previously created compensation image for scanning the label area The compensation image improves the effectiveness of barcode rea...

Page 30: ...ameters properly follow the instructions below By moving the mouse over the buttons the software displays information about the setting you are interested in Brightfield Compensation Figure 10 Brightfield Compensation Camera Selection This field displays the active camera Currently there is only one active camera for Brightfield mode Objective Selection The installed objective is selected for crea...

Page 31: ...in the box Use Selection Box Select this if you have a compensation slide that has only certain predefined areas for creating compensation images due to the type of the compensation slide or because the compensation slide has faded in some areas To create the compensation image s click Click the Stop button to interrupt the process The Progress bar at the bottom of the window displays the progress...

Page 32: ...12 Base settings Objective type You can select the type and the magnification level of the objectives Camera type Select the correct camera type in Pos 1 Camera type then the software automatically loads the camera settings upon clicking the button Brightfield ICC Profile The color managing scheme can be selected from the drop down menu Save settings by clicking the button October 9 2015 Rev 1 3DH...

Page 33: ...ESK 1 20 User s Guide 4 Software Description Filter Settings Figure 13 Filter settings The settings on this panel apply to Fluorescent mode therefore irrelevant and inactive October 9 2015 Rev 1 3DHISTECH Ltd 33 76 ...

Page 34: ...ra rotation Caution If you have transported the product or changed the camera there is a possibility that the camera does not display the images properly If you experience such problems contact service support Do not modify these settings yourself October 9 2015 Rev 1 3DHISTECH Ltd 34 76 ...

Page 35: ...cation call settings for the following cases Slide Scanning Finished Scanning Error Occurred For Slide Scanning Finished and Scanning Error Occurred you can define the external software by entering the file path in the Application box or by browsing for it by clicking If you define parameters in the Parameters box it will pass them to the software that it starts If no parameter is defined then the...

Page 36: ...er add or remove users and also set the message body in the Message subject field by entering a text for example to inform recipients on the location of the scanner or the slide set you are working with Click Test to test settings by sending an e mail to the address defined in the E mail To field after you have finished to check their relevancy before applying them First click Setup to display the...

Page 37: ...utton to send test e mail to the specified address There are 2 feedback messages that come up after sending the e mail An e mail has been sent to the recipient An error has been occurred The e mail could not be sent Check parameters then retry If you have properly set messenger settings click OK then click Users in the Message Call section of the Event Call Settings window to address functions to ...

Page 38: ...ettings You can also set profile settings here and start scanning in Brightfield Manual mode For more information on profiles see section 4 5 Using Profiles Figure 18 Main window of control software 1 Tab panel with scanner functions and settings 2 Scan camera live image display area 3 Preview image recorded by the preview camera 4 Navigation and Preview image editor toolbar 5 Information area abo...

Page 39: ...d that it rests perfectly horizontally on the stage To enter the slide name To enter the slide name manually type it into the Slide Name field To use the barcode as a slide name if a compatible barcode sticker is affixed on the slide click First the software automatically detects the barcode on the slide then the barcode string will be displayed in the Slide Name field You can also find the button...

Page 40: ...e current profile click Save To select an existing profile from the selection menu click Load To delete the current profile click Delete To access the Scan Options window click Note These settings are only applicable if you do not use profiles If you use profiles the settings stored in them are used 4 3 2 Scan Options window Figure 20 Scan options window October 9 2015 Rev 1 3DHISTECH Ltd 40 76 ...

Page 41: ...function is only available for products installed with Stingray camera It is not accessible for configurations installed with CIS camera Image fields will be saved with native pixel resolution If this function is disabled you will work with reduced pixel resolution These resolution differences are not really visible with the human eye what is important in this feature is the resulting smaller file...

Page 42: ...r application see section 4 4 Scan To Server If you enable the Directory function the virtual slides will be digitized in the specified folder You can browse for the folder with the Browse button If you specify a folder that does not exist the software will create it All parameter inputs are saved for the next software session 4 3 3 Preview tab The Preview tab offers options for threshold setting ...

Page 43: ...nsitive scanning but it is possible that some of the smaller specimen areas will not be digitized Scan area inside the marker If the specimen cannot be identified as required for digitization for example weak contrast it can help to encircle completely the specimen with a black marker pen You have to draw a full circle with the pen to enable the detection Marker pen detection will make specimen de...

Page 44: ...s Warning Do NOT use Remove coverslip function on TMA slides The layout of TMA slides may mislead the detecting algorithm Fill holes in scan area If selected the holes inside the specimen will also be scanned Often these holes contain small sample objects that otherwise would not be scanned Remove specks If selected the scanner does not digitize objects smaller than the predetermined size This hel...

Page 45: ...review mask or TMA grid Use selection box If selected only the area inside the box is scanned It is useful if you only want to scan the area that is equivalent with the size of the coverslip so external specks will not be scanned 4 3 4 Preview area scanned by preview camera The preview image shows the area that will be digitized with a red and or an orange mask If the Use selection box is enabled ...

Page 46: ...l not be scanned These areas are not digitized on the basis of their position for example they are outside the selection box They are detected by the selected threshold value Marker If Scan area inside the marker is selected in the Preview tab the preview camera detects the marker pen The inside areas appear in red color and will be digitized Areas scanned in This color is only displayed during sc...

Page 47: ...g with the mouse TMA tool To undo your last action click maximum of 40 changes To position the specimen holder that holds the current slide click Position sample and then click on the area of interest on the preview image This will define the Live image that you can access from the Focus panel To display the original image created by the preview camera click To shrink or enlarge the preview image ...

Page 48: ...n the following four tabs Select tab Figure 28 TMA tool Select To create a new TMA object click New To select an existing TMA object click on the object name The selected object name will highlight To delete an existing TMA object click Delete Block tab Figure 29 TMA tool Block Set the type of TMA block on this panel October 9 2015 Rev 1 3DHISTECH Ltd 48 76 ...

Page 49: ...t the appropriate values with the arrows 4 Enter the average diameter of the spots in the Diameter of spots µm field or select the appropriate value with the arrows Grid tab The Grid tab displays the previously set grid settings Figure 30 TMA tool Grid Move tab The Move tab allows the rotation and moving of the TMA selection spots that are displayed on the preview image To move the spots enter the...

Page 50: ...Pannoramic DESK 1 20 User s Guide 4 Software Description Figure 31 TMA tool Move tab with preview October 9 2015 Rev 1 3DHISTECH Ltd 50 76 ...

Page 51: ...ser s Guide 4 Software Description Label area To rotate the label area by 180 select the Rotate image by 180 degrees checkbox Figure 32 Label area 4 3 5 Focus tab Figure 33 Focus tab October 9 2015 Rev 1 3DHISTECH Ltd 51 76 ...

Page 52: ... as a traditional microscope Define the manual focus step size by adjusting the Step size slider Search for the sharpest focus on the current field of view automatically by clicking Multilayer mode Select Single layer option if you want to scan a single layer of the slide By choosing the Extended focus option the software selects the sharpest image from each focus level for each image field and co...

Page 53: ...b This tab contains control and maintenance functions Show images during scanning If you select this option you will be able to see the currently scanned image field during scanning in the live image display field This slows down digitization process and should be activated only with purpose October 9 2015 Rev 1 3DHISTECH Ltd 53 76 ...

Page 54: ...ive images The display field for live images displays images of the scan camera The live image is displayed if no current digitization process is running or Show images during scanning option is enabled Figure 36 Display field and toolbar for live images October 9 2015 Rev 1 3DHISTECH Ltd 54 76 ...

Page 55: ... were detected Calculates white balance only for color cameras Saves the current image as BMP or JPEG file Fast live mode live images are displayed with a frequency of at least 10 images per second If an exposure time is set to more than 100 ms the pixel values are multiplied accordingly for example if an exposure time of 200 ms is set the pixel values are multiplied by 2 and the exposure time is ...

Page 56: ... line Available disk space of the information area of Scan Progress You can interrupt a running digitization process by clicking This button is only enabled if a digitization process is in progress 4 4 Scan To Server The ScanToServer function is available for you to scan directly to a dedicated CaseCenter server To setup connection with a designated server follow the steps below 1 Click Scan Optio...

Page 57: ...n click Connect Figure 38 Scan to CaseCenter option 3 Type in or select the server address from the Connect to drop down list and give the User name and Password Activate the Secure connection option to enable HTTPS connection Click Connect when finished Figure 39 Server Log in window October 9 2015 Rev 1 3DHISTECH Ltd 57 76 ...

Page 58: ...he Scan Options window Slide settings in the Scan To Server window 1 In the Routine work tab panel click the button 2 The Scan To Server window opens Figure 41 Scan To Server window Slide data is displayed in a form Upon activating the Read barcode if no name is defined on the list option the slide barcode will be read and set in the Barcode field If you do not want to use this option you can type...

Page 59: ...f the barcode will be displayed and set as slide name 123 Case and Block fields can be set this way as well if necessary NOTE When filling slide data fields press TAB to move to the next cell Figure 42 Profile selection in the Scan To Server window Profile can be added by selecting a list entry from the Profile drop down list Select a folder where the slide will be scanned to After clicking the Fo...

Page 60: ...lly selected as scanning destination based on the CaseID containing slide and patient data stored on HIS LIS Therefore Slide name Folder Case and Block field data cannot be modified only a profile can be selected Figure 44 Scan To Server window during Barcode parsing Click Done when finished or Cancel to quit the Scan To Server window If you save the form slides will be scanned to server according...

Page 61: ...ON OFF available if Stingray F146C camera is used Determination of the area to be scanned Use of automatic threshold specimen Use of specimen threshold value Use of marker pen threshold and specimen threshold values Remove coverslip function ON OFF Use of scan everything within marked area function ON OFF Filling of holes ON OFF Use of removing specks according to the set value ON OFF Use of saved...

Page 62: ...ffects damaging to health Decontaminate infectious materials after use Dust specks and humidity may decrease or harm the operating capability of the device Protect it with dust cover when not in use Switch the power off before covering it Avoid touching the lamp bulb Wait for about 15 minutes to allow the bulb to cool down and disconnect the power plug before changing the bulbs Defective devices d...

Page 63: ...for the Pannoramic products If any problem arises that does not have solution in this manual contact the service personnel The product does not start Power supply is interrupted 1 Make sure that the 24V external power supply is on 2 Check cable connections Control software has stopped Name of currently selected slide for digitization already exists and is open in Pannoramic Viewer Close virtual sl...

Page 64: ...acing the halogen lamp Virtual slide is improperly focused almost across its entire surface Focus range enabled Disable Limiting focus range Poor image quality Imaging error for example chromatic aberrations Contact 3DHISTECH Service for support Warning Do not modify the camera driver settings as the default factory settings give the most accurate images Virtual slides show poor stitching in Panno...

Page 65: ...ded focusing option if not required Z stack option is activated Deactivate the Z stack option if not required Image is out of focus Focus range limit is activated Deactivate the Limiting focus range option Focus range setting is incorrect Change focus range setting All individual images show a shading effect Algorithm failed to find empty image fields Capture a reference image with a blank slide s...

Page 66: ... replace the halogen lamp or to perform other maintenance work and to interrupt a running digitization procedure because of slide jamming the protective cover of the Pannoramic DESK need to be removed in order to eliminate the problem Before removing the cover exit the control software switch off the 24 V power supply and disconnect the power cables Caution Only instructed persons or service techn...

Page 67: ...d maintenance or repair work reattach the cover Push it horizontally towards the basic unit 4 until the two protruding guide pins 6 fit into the corresponding recesses of the basic unit 4 Note that the rear part of the cover must be pushed over the top part of the blue basic unit 7 5 To lock the cover in place insert the screws again and fasten them Caution Make sure that no cables are squeezed as...

Page 68: ...he hole 5 on the heat sink body 7 Remove the lamp mount with the lamp 8 Remove the defective lamp 4 from the lamp mount 9 Insert a new bulb into its mount Warning Do not touch the bulb with your bare fingers Use a dry clean lint free cloth or gloves when inserting the bulb into the mount 10 Taking care that the two clamps 3 on the lamp mount are correctly aligned to the hole 5 on the heat sink bod...

Page 69: ...d slide cancel the running digitization process terminate the software session and turn the power off Any manipulation performed on a mechanical component must be followed by a complete device initialization routine software restart Figure 48 Removing a slide 1 Exit control software 2 Switch off power supply and disconnect power cable 3 Remove the cover see section 5 3 1 Removing the protective co...

Page 70: ...r 8 Connect the power cable of the power supply to the power outlet and switch on the power supply 9 Launch control software 5 3 4 Transporting Pannoramic DESK The following procedure shows how to transport Pannoramic DESK Caution Do not lift or transport Pannoramic DESK by holding its protective covers Note The base unit does not have carrying handles for transporting One or two persons can carry...

Page 71: ...SK do the following 1 Turn power supply off and detach all cables 2 Transport Pannoramic DESK 3 After finishing transportation reestablish cable connections Caution Readjust camera angle if necessary after transporting and reinstalling the Pannoramic DESK October 9 2015 Rev 1 3DHISTECH Ltd 71 76 ...

Page 72: ...ply approx 550 g Ambient conditions Transportation shipment packed Permissible ambient temperature 40 C to 70 C Storage Permissible ambient temperature 0 C to 60 C Permissible relative air humidity non condensing 75 Operation Permissible ambient temperature 10 C to 35 C Permissible relative air humidity 75 Altitude at installation site max 2000 m Atmospheric pressure 500 hPa to 1060 hPa Degree of ...

Page 73: ...ower consumption of the complete device 75 VA 24V external mains power supply unit Intended site closed room facilities Electrical protection class I Internal protection degree IP 20 Electrical safety in compliance with UL 2601 1 TÜV EN 60601 CE EN60601 1 2 2001 Over voltage category II Radio interference suppression in accordance with CISPR 11 EN 55011 class B Line input voltage 100 V AC to 240 V...

Page 74: ...g UPC A Numeric encoding DataMatrix Numeric encoding Alpha encoding AlphanumericPunc encoding AlphaNumeric encoding ASCII encoding IS08 encoding Maxicode Encoding mode 2 3 4 5 6 MicroPDF417 Standard encoding type PDF417 Standard encoding type QR QR code Model 1 2 encoding Scan cameras Stingray F146C IRF Medical camera 1 2 type progressive scan CCD Pixel size 4 65µm x 4 65µm Camera resolution 1388 ...

Page 75: ...on with 20x objective and 1x C mount adapter 0 24µm Connection type Dual CameraLink full 10 taps Control computer minimum system requirements Operating System Microsoft Windows 7 SP1 64 bit EN CPU Intel Core i5 4590 RAM 8 GB DDR3 1600 Hard disk 500 GB Optical drive DVD RW Display 96 dpi Graphics SXGA 1280 1024 True color Input devices keyboard mouse Ports 2 x Fire Wire 1394 A 1394 B 4 x USB 2 0 GB...

Page 76: ...47 51 52 53 61 64 65 H Halogen lamp 64 66 67 73 M Marker 43 46 61 P Pannoramic 6 Pannoramic DESK 6 8 9 10 12 13 15 17 19 23 25 26 63 66 70 71 Pannoramic Viewer 18 52 63 64 Preview 20 22 24 26 31 38 42 43 44 45 46 47 49 50 52 74 R RGB 55 S Scan 12 14 17 31 38 39 40 41 42 43 44 45 46 52 53 54 56 61 64 70 74 Scan To CaseCenter 42 Scan To Server 58 59 60 Scanner 6 31 44 ScanToServer 42 56 Stingray 16 ...