Guidance and manufacturer's declaration - electromagnetic immunity
The device is intended for use in the electromagnetic environment specified
below. The user of the device should make sure that it is used in such an
environment.
Immunity Test
IEC 60601 Test
Level
Compliance
Level
Electromagnetic
Environment - Guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±8 kV contact
±2 kV, ±4 kV,
±8 kV,±15 kV
air
±8 kV contact
±2 kV, ±4 kV,
±8 kV,
±15 kV air
Floor should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should
be at least 30%.
Electrical
fast
transient / burst
IEC 61000-4-4
±2 kV for
power supply
lines
±2 kV for
power supply
lines
Mains
power
quality
should be that of a typical
commercial or hospital
environment.
Surge
IEC 61000-4-5
±1 kV
line(s) to line(s)
±1 kV
line(s) to line(s)
Mains
power
quality
should be that of a typical
commercial or hospital
environment.
Voltage
dips,
short
interruptions
and
voltage
variations
on
power supply
input lines
IEC 61000-4-11
0%
U
T
; 0.5
cycle
At 0°, 45°, 90°,
135°, 180°,
225°, 270° and
315°
0%
U
T
; 1 cycle
70%
U
T
; 25 / 30
cycle
At 0°
0%
U
T
; 250 /
300 cycle
0%
U
T
; 0.5
cycle
At 0°, 45°, 90°,
135°, 180°,
225°, 270° and
315°
0%
U
T
; 1 cycle
70%
U
T
; 25 / 30
cycle
At 0°
0%
U
T
; 250 /
300 cycle
Mains
power
quality
should be that of a typical
commercial or hospital
environment. If the user of
the
device
requires
continued
operation
during
power
mains
interruptions,
it
is
recommended that the
device be powered from
an uninterruptible power
supply or a battery.
Power
frequency (50 /
60
Hz)
magnetic field
IEC 61000-4-8
30 A/m
30 A/m
Power
frequency
magnetic fields should be
at levels characteristic of
a typical location in a
typical commercial or
hospital environment.
Note:
U
T
is the AC mains voltage prior to application of the test level.