2. Description 

The  Hall  effect  basic  apparatus  provides  con-
tacts  and  power  supply  for  the  circuit  boards 
with  a  germanium  crystal  (1008522,  1009810 
and  1009760),  also  holding  them  in  place  in  ex-
periments  measuring  the  Hall  effect  and  how  it 
depends  on  temperature,  magnetic  field  or  sam-
ple  current,  as  well  as  experiments  on  electrical 
conductivity.  

The  apparatus  contains  an  adjustable  constant 
current  source  for  the  sample  current,  a  meas-
urement  amplifier  with  offset  compensation  for 
the  Hall  voltage  and  adjustable  heating  for  sam-
ples. The heating automatically cuts out at 170°C 
to  protect  fragile  germanium  crystals.  Hall  volt-
age,  sample  current,  sample  voltage  and  tem-
perature  can  all  be  read  off  from  the  display, 
which  can  be  switched  between  these  various 
modes. In addition, Hall voltage and sample volt-
age  can  be  tapped  directly  via  contacts  on  the 
front,  and  three  voltage  equivalents  for  Hall  volt-
age or sample voltage, sample current and sam-
ple  temperature  on  the  side.  On  the  right-hand 
side of the case there are three mini-DIN sockets 
for  computer-based  measurement  of  the  quanti-
ties mentioned above using a 3B NET

log

TM

 unit. 

The equipment is set up on the U-shaped core of 
the multi-piece transformer kit. The magnetic field 
can  be  measured  using  a  magnetic  field  sensor 
adjacent to the crystal. 

 
 

3. Contents 

1 Hall effect basic apparatus 

1 U-shaped holder 

2 Connector leads with 8-pin mini-DIN plugs 

Instruction manual 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

4. Controls 

4.1  Decoupled  measurement  outputs  for 

measuring voltage equivalents for quan-
tities 

       

                  

 

Fig.1  Measurement  outputs  using  4-mm  sockets  and 
8-pin mini-DIN sockets for 3B NET

log

TM

 

a/e 

Hall voltage/sample voltage* 

b/f 

Sample current 

c/g 

Sample temperature 

Earth socket 

*  only  if 

U

P

  is  being  displayed,  otherwise  it  out-

puts the Hall voltage 

 

4.2  Measurement  outputs  for  direct  meas-

urement 

 

Fig. 2 Measurement outputs for direct measurement 

h1, h2  Sample voltage 
i1, i2 

Hall voltage 

 

Caution:  do  not  any  apply  an  external  volt-
age to the measurement outputs! 

 

4.3  Buttons  for  switching  display  and  corre-

sponding LEDs 

 

Fig. 3 Buttons for switching display 

Hall  voltage or sample voltage (Hold down 
for 2 s to display sample voltage) 

Sample current 

Sample temperature 

Summary of Contents for Hall Effect Basic Apparatus

Page 1: ...equipment conforms to safety regulations for electrical measuring instruments and control and laboratory equipment as per DIN EN 61010 part 1 and is designed to conform to protection class 3 It is intended for use in dry rooms which are suitable for the use of electrical equipment If the equipment can no longer be operated without hazards arising it should immediate ly be withdrawn from use Do not...

Page 2: ...and side of the case there are three mini DIN sockets for computer based measurement of the quanti ties mentioned above using a 3B NETlogTM unit The equipment is set up on the U shaped core of the multi piece transformer kit The magnetic field can be measured using a magnetic field sensor adjacent to the crystal 3 Contents 1 Hall effect basic apparatus 1 U shaped holder 2 Connector leads with 8 pi...

Page 3: ...on circuit board 1009810 or 1 N doped germanium on circuit board 1009760 or 1 Undoped germanium on circuit board 1008522 1 U shaped core 1000979 1 Pair of pole pieces for Hall effect plus clamps 1009935 2 Coils 600 windings 1000988 1 Magnetic field sensor 2000 mT 1009941 Place the U shaped core on a stable level surface Place coils over the core in such a way that their connectors are facing the f...

Page 4: ...ply and set an output voltage of 12 V Connect the coils to the DC power supply Connect the magnetic field sensor to the 3B NETlog unit Select sample current IP for max sample current see instruction manual for germani um crystal select the Hall voltage on the console and adjust the zero point for the Hall voltage with the help of the compensation trimmer 7 1 1 Hall voltage as a function of sample ...

Page 5: ...A 230 1003316 or 1 Transformer with rectifier 3 A 115 V 1003315 Select a small sample current IP Since the current will cause the sample to heat up you should not exceed a current of 5mA Switch the display to sample voltage Turn on the heating controller and record the sample voltage UP as a function of tempera ture For your measurement plots see the manu al for the relevant circuit boards 8 Care ...

Page 6: ......

Reviews: