3B SCIENTIFIC PHYSICS PS 400 Instruction Manual Download Page 2

 

1. Safety instructions  

Plug-in power supply 230 V (1001014)/ 115 V 
(1009545): 

 

Operation only with the mains voltage ac-
cording to the imprint. 

 

Do not use the device if there is visible 
damage to the housing, plug or cable. 

PS 400: 

 

Connect or disconnect the control unit to 
the sliding rail inly without connected 
power supply. 

 

Adjust the clutch force with the screw (14) 
(allows the belt to slide in case of collision 
at the end of the track or on sensitive com-
ponents, see 5.). 

 

The sensor cable 1021514 is needed for 
the CMA interface (8). 

 

Do not apply voltage to the voltage output 
V

out

 (5).

 

  

2. Description  

The  Remote-Controlled  Positioning  System 
PS 400  enables  fast  and  exactly  reproducible 
positioning of components in experiments. 
Fixed and programmable positioning steps with 
assigned and programmable moving velocities 
offer a high degree of flexibility with particularly 
simple operation. 
Experiments  can  be  performed  quickly  and 
without disturbing influences of the person per-
forming the experiment. 
The  CMA  interface  (8)  enables  recording  the 
position of the sled on the sliding rail via com-
puter or VinciLab (1021477). The position of the 
sled can also be determined via the voltage out-
put (5). Measurement data from other sensors 
can thus be recorded as a function of the sled's 
position and quickly evaluated. 

3. Scope of delivery  

1  Sliding rail 
1  Control unit PS400 
1  Stand rod: Ø = 10 mm, length: 200 mm 
1  Tripod muff 
1  Plug-in  power  supply  230 V  (1001014)  or 

115 V (1009545) 

 

 

4. Technical data 

Sliding rail: 

Weight: 

2 kg 

Moving distance:  

405 mm 

Stand rod mount:  

Ø = 10mm 

Dimensions: 

580x120x50 mm

 

Control Unit PS400: 

The control unit has fixed programmed keys 

step size

 (3). Pressing a button once moves 

the sled the distance 

s

 at the velocity 

v

:  

"1“ 

s

 = 1 mm 

v

 = 1 mm/s 

"5“ 

s

 = 5 mm 

v

 = 5 mm/s 

"10“ 

s

 = 10 mm 

v

 = 10 mm/s 

"50“ 

s

 = 50 mm 

v

 = 25 mm/s 

"100” 

s

 = 100 mm 

v

 = 50 mm/s 

The "100" key is also programmable. The dis-
tance can be selected from 1 mm to 405 mm, 
the velocity from 1 mm/s to 100 mm/s (see 5.). 

The movement of the sled is uniform: 

v

 = con-

stant. 

Precision of moved distance: ± 0.1 % 

Connetions: 

1x  Output voltage (5),  

V

OUT

 = 0 V bis ± 4,05V 

1x  CMA interface (8) 

Sensor cable 1021514 required 

1x  Sub-D 

Sliding Rail

 (1) 

1x  Hollow bushing (4), supply voltage 
2x  Ground socket (7) 

Supply voltage: 

12 V AC 

Dimensions: 

110x170x30 mm3 

Weight: 

1 kg 

5. Operation 

  Connect the sliding rail and the control unit. 

After connect the power supply to the con-
trol unit. The current position of the sled is 
the so called "home" position (voltage U

out

 

between (5) and the reference point (7) is 
0 V). 

  By  pressing  the  "home"  key,  the  sled 

moves  back  to  the  "home"  position  from 
any position. 

  A new "home"  position  can  be set  at  any 

desired  sled  position  by  pressing  the  two 
function keys (6) simultaneously.  

Reviews: