5. Current measurements 

 

The Escola 30 is 

not permitted

 to be 

used  for  measurements  on  low-volt-
age  mains  distribution  equipment, 
such as mains sockets, circuit breakers 
and fuses etc. 

 

The  nominal  voltage  of  the  electricity 
source must not exceed 30 V and its 
nominal  current  must  be  no  higher 
than 3 A. 

 

The multimeter is to be connected 
in series with the load on the side 
where the voltage  with respect to 
ground is lowest! 

 

Turn off the circuit before the Es-
cola 30 meter is connected into it. 

 
 

 

 
 

 

Connect  the  side  with  the  lower  potential  to 
the left-hand socket. 

 

Connect  the  common/ground  lead  first  and 
only then the signal lead. 

 

5.1  DC currents (mA=): 

 

Set the rotary switch to the desired DC current 
range.  If  measurements  are  made  when  the 
magnitude of the variable is unknown, always 
select a large measuring range before shifting 
down to lower ones. 

 

Set  the  three-way  switch  to 

  or,  if  the 

zero point is to be set in the centre, to

 

5.2  AC currents (mA~): 

 

Set the rotary switch to the desired AC current 
range.  If  measurements  are  made  when  the 
magnitude of the variable is unknown, always 
select a large measuring range before shifting 
down to lower ones. 

 

Set the three-way switch to 

 
 

6. Voltage measurements 

 

The Escola 30 is 

not permitted

 to be 

used  for  measurements  on  low-volt-
age  mains  distribution  equipment, 
such as mains sockets, circuit breakers 
and fuses etc. 

 

The  nominal  voltage  of  the  voltage 
source must not exceed 30 V. 

 
 

 

 
 

 

Connect  the  side  with  the  lower  potential  to 
the left-hand socket. 

 

Connect  the  common/ground  lead  first  and 
only then the signal lead. 

 

6.1  DC voltages (V=): 

 

Set the rotary switch to the desired DC voltage 
range.If  measurements  are  made  when  the 
magnitude of the variable is unknown, always 
select a large measuring range before shifting 
down to lower ones. 

 

Set  the  three-way  switch  to 

  or,  if  the 

zero point is to be set in the centre, to

 

6.2  AC voltages (V~): 

 

Set the rotary switch to the desired AC voltage 
range.If  measurements  are  made  when  the 
magnitude of the variable is unknown, always 
select a large measuring range before shifting 
down to lower ones. 

 

Set the three-way switch to 

 
 
 

Symbols for measuring ranges 

mA=  DC (unit mA) 

mA~  AC (unit mA) 

V= 

DC voltage (unit V) 

V~ 

DC voltage (unit V) 

 
 
 

~

~

~

~

~

~

~

~

Summary of Contents for Escola 30

Page 1: ...tities with frequen cies of up to 40 kHz can be measured All the pos sible measuring ranges can be selected by means of a single rotary switch The excess current protection of the Escola 30 au tomatically limits the power when overloaded Af ter a brief cooling phase the multimeter au toma tically switches on again The use of a robust moving coil galvanometer and a mechanically resilient casing mak...

Page 2: ...cting the battery test function When measuring current make sure the elec tricity is turned off before the analogue multi meter is connected into the circuit When making measurements always con nect the ground lead first Disconnect the sig nal measurement lead before unplugging the ground Turn off the multimeter before opening the casing disconnect the power to the circuit and the measuring leads ...

Page 3: ...range If measurements are made when the magnitude of the variable is unknown always select a large measuring range before shifting down to lower ones Set the three way switch to 6 Voltage measurements The Escola 30 is not permitted to be used for measurements on low volt age mains distribution equipment such as mains sockets circuit breakers and fuses etc The nominal voltage of the voltage source ...

Page 4: ...ckets Power supply Battery 1x 1 5 V AA IEC LR6 Automatic cut off after 45 min 10 min Electromagnetic compatibility Interference emission EN 55011 2009 Interference resistance EN 61326 1 2013 Operating conditions Ambient temperature 5 C 23 C 40 C Storage temperature 20 70 C Relative humidity 85 with no condensa tion General data Shock test max 147 m s Dimensions 100x150x50 mm3 ap prox Weight 300 g ...

Page 5: ...e a soft cloth slightly soaked in alcohol or a paint brush Dirt or moisture in the measurement sockets can affect readings Shake out any dirt that may be in the meas urement sockets Soak a new swab with isopropyl alcohol and work around the inside of each measurement socket 11 Disposal The packaging should be disposed of at local recycling points Should you need to dis pose of the equipment it sel...

Page 6: ......

Reviews: