Siesta Nasal  Mask 

User Manual 

Components of the Mask 

Soft sleeves� 

Frame 

� 

Clips 

Cushion assembly 

Components of the Siesta Nasal Mask 

Siesta Nasal Mask 

Thank you for choosing 3B Medical's Siesta Nasal Mask. 

The Siesta Nasal  Mask is designed to minimize contact with your face, 

thus ensuring that you feel comfortable during therapy. This user manual 

provides  you with the information you  need  for the correct use of your 

mask. 

The Mask is not made with natural rubber latex. 

Intended Use 

The  Siesta  Nasal  Mask  channels  airflow  non-invasively  to  the  patient 

from  a  positive  airway  pressure  device  such  as  a  continuous  positive 

airway pressure (CPAP) or bi-level system. 

The Siesta Nasal Mask is: 

To  be  used  by  adult  patients  (>  66  lb  /  30  kg)  for  whom  positive

airway pressure therapy has been prescribed.

Intended for single-patient reuse.

CAUTION: In the US,  Federal law restricts the sale of this device by or 

on the order of 

physician. 

Before Using the Mask 

.&_ 

WARNINGS

The vent holes must be kept clear.

This  mask  should  only  be  used  with  CPAP  or  bilevel  devices 

recommended by a physician or respiratory therapist. 

The mask should not be used unless the device is turned on.  Once 

the mask is fitted, ensure the device is blowing air.

Explanation: 

CPAP  and  bilevel  devices  are  intended  to  be  used 

with  special masks  (or connectors)  which have  vent holes to allow 

continuous flow of air out of the mask. When the device is turned on

and  functioning  properly,  new  air  from  the  device  flushes  the 

exhaled  air  out  through  the  mask  vent  holes.  However,  when  the

device is not operating, insufficient fresh air will be provided through

the mask,  and  the  exhaled  air  may  be  rebreathed.  Rebreathing  of 

exhaled  air  for  longer  than  several  minutes  can,  in  some 

circumstances,  lead  to  suffocation.  This  applies  to  most  models  of 

CPAP or bilevel devices. 

This mask is made with silicone parts.  Please do not use if patient

has silicone allergy or sensativity. 

This mask  should not be used on patients who are  uncooperative, 

obtunded, unresponsive, or unable to remove the mask.

Follow all precautions when using supplemental oxygen. 

Oxygen flow must be turned off when the CPAP or bilevel device is 

not  operating,  so  that  unused  oxygen  does  not  accumulate  within

the device enclosure and create a risk of fire.

Oxygen  supports  combustion.  Oxygen  must  not  be  used  while

smoking  or in  the  presence  of  an open  flame.  Only  use oxygen  in

well ventilated rooms.

At a fixed flow rate of supplemental oxygen flow, the inhaled oxygen

concentration  varies,  depending  on  the  pressure  settings,  patient 

breathing pattern, mask, point of application and leak rate. 

The  technical  specifications  of  the  mask  are  provided  for  your

clinician  to  check  if  it  is  compatible  with  the  flow  generator.  If  it  is 

used  beyond  technical  specifications  or  used  with  incompatible 

devices,  the  seal  and  comfort  of  the  mask  may  not  be  effective, 

optimum therapy may not be achieved,  and leak,  or variation in the 

rate of leak, may affect the function of the flow generator. 

Stop  using  the  Siesta  Nasal  Mask  and  consult  your  physician  or

sleep therapist,  if you have ANY adverse reaction to the use of the 

mask.

Refer to your CPAP or bilevel device manual for details on settings 

and operational information.

Remove all packaging before using the mask. 

Fitting the Mask 

Attach the headgear 

straps by threading them 

through the gaps of 

forehead frame and clips. 

Adjust the upper headgear 

straps evenly until they are 

comfortable and fsecure, 

and reattach the mask. 

Hold the mask over your 

nose and pull the 

headgear over your head. 

Adjust the lower headgear 

straps evenly until they are 

comfortable and fsecure, 

and reattach the mask. 

Disassembling the Mask 

Bringing the lower 

headgear straps below 

your ears,  connect the 

clips to the Frame 

assembly. 

Mask is worn as shown. 

Undo and remove the upper headgear 

Remove the Cushion assembly from 

straps out of the slots on the Frame. 

the Frame assembly. 

Remove the Short tube from the Frame 

assembly. 

Cleaning the Mask 

Notes: 

The mask and headgear can only be cleaned through hand washing

or placed in the Lumin device 

The headgear may be washed without removing the clips.

When  washing  the  mask,  the  Elbow  and  the  Frame  cannot  be

disassembled from the Frame assembly.