12
TECHNICAL SPECIFICATIONS
Electric two-cylinder piston compressor with lifetime lubrication. Metal casing with epoxy powder coating
Risk class under Directive 93/42/EEC:
IIa
Suction class:
High vacuum / Low flow
Voltage:
12VDC 45 W
Power consumption:
3.75A
Adjustable vacuum level:
from 0 to -0.85 bar (-85 KPa)
Vacuum gauge precision class:
2.5% according to UNI EN 837
Air flow rate:
28 L/min operating without restri/- 10%
Temporary use:
maximum 45 minutes
Dimensions:
37 (length) x 10 (width) x 24 (height) cm
Weight:
5.8 kg approx.
Noise level:
58 dBA approx.
Multi-voltage switching battery charger:
PRI: 100-240V~ 50-60Hz
SEC: 14V DC 4.29 A
Internal battery:
12VDC 4.5 Ah lead, hermetic
Battery charge duration:
45 minutes at maximum suction power
Cigarette lighter plug cord fuse:
F6.3 A-250 V, delayed; dimensions Ø 6.3 x 30 mm
Operating conditions: Temperature: min. 0°C max. 40°C – Air humidity: min 10% maximum 95%
Storage conditions: Temperature: min. -10°C max. 50°C – Air humidity: min. 10% max. 95%
Atmospheric pressure for operation/storage: min. 690 hPa - max. 1060 hPa
SYMBOLS USED
Type BF device
Class II device
It is compulsory to carefully read the instructions before using this device
Switch on
Switch off (or battery on charge)
Direct current
Alternating current
Never use the device when taking a bath or a shower
The device contains a sealed lead-acid battery. It must be disposed of in accordance with current regulations on the
disposal of toxic-harmful waste.
IP21
The device’s casing is protected against solids 12.5 mm or more in diameter, against vertically falling drops of water and against
finger access to hazardous parts.
Ethylene oxide sterilisation
Disposable
3A HEALTH CARE S.r.l.
Via Marziale Cerutti, 90F/G
25017 Lonato del Garda (BS) - ITALY
Compliant with Medical Devices Directive 93/42/EEC