3A HEALTH CARE MAXIASPEED Instruction Manual Download Page 5

10

•  In view of their length, the power cord and connection hose could constitute a strangulation hazard.
•  The device is equipped with a safety fuse, easily inspected in the event of a fault, located on the back of the device in the panel 

socket. 

Pull out the power supply plug before inspection

.

•  The antibacteria filter is a 

disposable

 element which must be replaced after every application.

•  The device comes into contact with the patient via the single-use cannula

 (NOT supplied with the device)

: any aspiration 

cannulas, purchased separately from the device, must comply with the requirements of the ISO 10993-1 standard on the bio-
compatibility of materials.

•  Do not lift the device using the transportation handle in any circumstance.

•  Before each use, follow the cleaning and disinfection operations as shown in the “

CLEANING AND DISINFECTION OPERA-

TIONS

” paragraph of this instruction manual.

INSTRUCTIONS FOR USE 

Before each use, ensure that all accessories are perfectly clean according to the instructions in the “

CLEANING AND DISINFECTION 

OPERATIONS

” section. 

1. 

Connect the device as shown in figure 1 (page 2):

1.1

 Connect the antibacterial filter (

2.

) to the “INLET” connector of the aspirator, ensuring that the side marked “

IN

” is facing 

towards the connector. 

Do not use the aspirator without the antibacterial filter, because from a bacteriological point 

of view, it becomes dangerous for the patient.

1.2 

Connect one end of the short silicone tube to the hose connection of the antibacterial filter and the other end to the 

“VACUUM” connection (4.1) of the vessel. The collection vessel is complete with overflow valve, lid and vessel in a clear mate-
rial (polycarbonate). The antibacterial filter also protects the suction circuit from possible contaminants sucked in during use.

1.3 

Connect the long tube to the “PATIENT” connector (4.2) of the selected plug.

2.  Operation of the telescopic handle

: for space reasons the aspirator is supplied with the telescopic handle lowered; to 

raise it, pull it up until the travel end hooks engage. To put the telescopic handle down again, press the travel end hooks on 
the external sides of the telescopic tubes with the tool provided (Fig. 2) and push the handle downwards until it closes on the 
aspirator casing.

3. 

Using the vacuum regulator (1.4) it is possible to set the desired vacuum value (bar). By turning the knob towards “+” a higher 
vacuum is obtained, while turning it towards “-“ a lower vacuum is obtained: these values can be read on the vacuum gauge 
(1.2).

4. 

Start the device by switching the main switch to the “I” (ON) position (1.5).

Version with pedal switch:

 If the device is equipped with the pedal switch, to use it connect the plug on the cable to the 

socket provided on the panel (1.10) on the back of the device and turn the switch (1.6) next to the main switch (1.5) to “I” 
(ON), allowing the device to be switched on/off using the pedal switch.

5.  

After the application, switch of the device, remove the power supply cable from the current socket and perform the cleaning 
operations as illustrated in the “

CLEANING AND DISINFECTION OPERATIONS

” paragraph.

6.  

There is a very capacious compartment (1.9) on the back of the device for putting the power supply cable back in after the ap-
plication, and if the device is equipped with the pedal switch, also for this latter.

FLOW DIRECTION

IN

IN

Vessel

Connection

Device

Connection

Summary of Contents for MAXIASPEED

Page 1: ...1 MAXIASPEED ...

Page 2: ...LOTTEUR CAGE SCHWIMMER KORB FLOTADOR JAULA 1 5 1 6 1 2 1 3 1 10 1 9 1 8 Fig Abb 1a Fig Abb 1b Fig Abb 3a Fig Abb 3b Fig Abb 5 OK KO GABBIA IN ASSE CAGE IN AXIS CAGE DANS L AXE FLUCHTENDER KORB JAULA EN EJE GABBIA FUORI ASSE CAGE OUT OF AXIS CAGE HORS AXE NICHT FLUCHTENDER KORB JAULA FUERA DE EJE 1 4 2 1 6 ...

Page 3: ...e prise appropriée 1 7 Roulettes antistatiques avec frein 1 8 Prise d alimentation avec fusible pouvant être inspecté 1 9 Compartiment porte câble et interrupteur à pédale 1 10 Prise pour interrupteur à pédale quand il existe 2 Filtre antibactérien 3 Tuyau court Ø 8x14 mm 4 Récipients 2000cc or 4000cc 4 1 Prise VACUUM 4 2 Prise PATIENT 5 Tuyau long Ø 8x14 mm 6 Raccord d aspiration 7 Outil pour poi...

Page 4: ...s for use The warnings and cautions described in the instruction manual must be observed IMPORTANT WARNINGS This is a medical device and it must be used by authorised personnel It must be operated as instructed in this use manual It is important for the operator to read and understand the use and maintenance information of the device Contact your retailer for any questions you may have MICROBIAL C...

Page 5: ...on vessel is complete with overflow valve lid and vessel in a clear mate rial polycarbonate The antibacterial filter also protects the suction circuit from possible contaminants sucked in during use 1 3 Connect the long tube to the PATIENT connector 4 2 of the selected plug 2 Operation of the telescopic handle for space reasons the aspirator is supplied with the telescopic handle lowered to raise ...

Page 6: ... cage is positioned correctly inside its housing It must be in axis with the plug aspiration hole so that the floating valve can block it in the circumstance in which the aspirated liquid exceeds the maximum volume the vessel can receive Do not connect the cage as shown in figure 3b it must be connected as shown in figure 3a Check that the float is fitted in the correct position and that it is fre...

Page 7: ... conditions Class of risk according to the 93 42 EEC directive Storage conditions Operating storage air pressure min 690 hPa max 1060 hPa Double pump version High flow high vacuum device 230V 50 60Hz 330VA approx 90 l min approx 50 dBA 1 m T2 5A 250V IIa Temperature min 10 C max 40 C Humidity min 10 max 95 Temperature min 25 C max 70 C Humidity min 10 max 95 approx 12 kg approx 13 kg approx 16 kg ...

Page 8: ...13 ...

Page 9: ...29 ...

Page 10: ...30 ...

Page 11: ...31 ...

Page 12: ... in your area or returning it to the distributor when buying a new device of the same type to be used with the same functions This procedure of separated collection of electric and electronic devices is carried out forecasting a European environmental policy aiming at safeguarding protecting and improving environment quality as well as avoiding potential effects on human health due to the presence...