7
CG000396 | Rev C
User Guide | Chromium X Series (X/iX) | Introduction
Product
Identification Label
Parameter
Specifications
Weight
41.4 lb (18.8 kg)
Instrument Dimensions with:
Tray close
Tray open
W D H
12.2”[31.0 cm] × 18.6”[47.2 cm] × 10.2”[26.0 cm]
12.2”[31.0 cm] × 22.8”[58.0 cm] × 10.2”[26.0 cm]
Electrical Requirements
Nominal
from a standard 3-prong wall
receptacle that includes a
safety ground pin
100-240 VAC, 50/60 Hz, 250 W
90-264 V operational range (+/- 10% of nominal)
Overvoltage Category II (standard receptacle)
Pollution
Degree 2 (Indoor Use Only)
Ventilation Requirement
Minimum 4” [10 cm]
Around all sides
Operating Temperature
64-82°F [18-28°C]
Use in a typical indoor laboratory environment. Extreme
temperature conditions will affect the sensitive reagents
used with the instrument.
Humidity
30-80% R.H. non-condensing
Altitude
Altitude up to 7500 ft [2286 m] above sea level
Power Cable Length
Power cables will be in
accordance with regional
specifications
6-7 ft [2 m] Standard
10.76”
12.2”
18.6”
Image of label for reference only
Product
Specifications
Summary of Contents for Chromium X Series
Page 5: ...Introduction 6 Introduction 7 Product Identification 7 Product Specifications ...
Page 8: ...Safety Compliance Information 9 Safety 10 Regulatory ...
Page 11: ...System Components 12 System Components 13 Chromium X iX Accessory Kits ...
Page 25: ...Menu Options 26 Menu Options 28 Firmware Update Procedure ...
Page 29: ...Maintenance 30 System Maintenance 30 Moving Shipping the Instrument ...