background image

FV454 Rev C   1/28/13
Page 4

8.)

 The ZEG urinal valve comes preset for both flow volume

and sensing distance.   Each valve is operated at the fac-
tory using water to insure proper function before being
packed for shipment.   The valve does not require water
pressure regulation for variation in water pressure within
the specified pressure range (20 psi to 120 psi) due to
internal pressure regulation.
Open the stop valve to wide open position using the ad-
justing screw 

(S)

 on the front of the stop valve.  The water

volume will never have to be adjusted with the stop valve
as on some conventional urinal installations.
Install vandal resistant stop cover to complete the installa-

tion.

APPENDIX A

TO CHANGE ACTIVATION DISTANCE FOR
THE OBJECT LOCK SENSOR

The valve does not have to be disabled in any way to change
the detection range.  Included with each valve is a Zurn “Magic
Magnet” that is used to initiate the auto-calibration mode.
Calibration is accomplished as follows.

1. Place the Zurn “Magic Magnet” 

(N)

 on the front of the valve

as shown in Figure 11 and moved around slightly until the
LED comes on.
2. The LED will blink once brightly, then nine times dimly, then
a final bright flash.  A solid red light will appear in the sensor
eye.  This means the electronics are ready to calibrate.
3. Remove the magnet and stand in front of sensor about 20”
away for the rest of the calibration.  The calibration process
takes about 60 seconds.  Do not move during this process.
4. When calibration is completed the solid red light will dis-
appear and the LED will blink twice brightly.

8

9.)

   LOW BATTERY WARNING

    The LED will double flash every 30 seconds when the bat-

teries reach a predetermined depletion level.  There are
approximately five hundred more flushes available once
the LED starts flashing, at which time the valve will stop
operating.   The LED will continue the flash sequence until
the batteries are replaced.

10

10.)

   FILTER

     

When accessing filter, be certain the stop valve is closed

at adjusting screw (S) Figure 8

.

     The filter is in the end of the tailpiece (see Figure 10). To

remove it for cleaning or inspection, loosen the locknut
completely 

(D)

 and loosen tube nut 

(H)

 one turn.  Swivel

the valve and remove the filter. It can be cleaned using
water and a brush. When reinstalling, the filter should snap
into the tailpiece securely.

H

S

N

11

M

D

4

Care and maintenance

Do not pressure wash any flush valve containing electron-
ics.  Water and soap blown through seals at high pressure
will damage electronics.  Clean valves only with mild
antibacterial soap and water using a soft cloth to clean and
dry the exterior.  Do not use cleaners containing abrasives
or chlorine compounds (especially ammonium chloride) to
clean flush valves beacause they will damage the chrome
plated and plastic (lens) surfaces.

Содержание ZEG6003EV

Страница 1: ...charges or any loss injury or damages whatsoever including incidental or consequential damages The sole and exclusive remedy shall be limited to the replacement of the defective goods Before installation and use the purchasershalldeterminethesuitabilityoftheproductforhisintendeduseandthepurchaserassumesallriskandliability whateverinconnectiontherewith Where permitted by law the implied warranty of...

Страница 2: ... stops are connected to the water supply and water pressure is available it is recommended that the supply piping be flushed to remove dirt metal chips etc from system A Before the valve is installed open each stop fully for a brief time and catch the water in a two gallon or larger bucket For multiple installations start with the stop valve closest to the water supply and work toward the most rem...

Страница 3: ...tall the AA cell batteries provided Remove the battery box cover screw using the 1 8 allen wrench provided Turn the top assembly upside down and the screw will drop down and can be used to pull the cover off Load the four heavy duty AA alkaline batteries following the battery orientation guide on the battery box cover Note that the coil springs always contact the flat negative end of the battery I...

Страница 4: ... the magnet and stand in front of sensor about 20 away for the rest of the calibration The calibration process takes about 60 seconds Do not move during this process 4 When calibration is completed the solid red light will dis appear and the LED will blink twice brightly 8 9 LOW BATTERY WARNING The LED will double flash every 30 seconds when the bat teries reach a predetermined depletion level The...

Страница 5: ...e detected Run through reset procedure below Hold hand close to lens to see if it will produce the 4 flash 2 flash pattern Close stop disconnect valve crack stop to check for water Close stop remove and inspect filter Check stop screw Check solenoid leads and insertion of solenoid plug Remove solenoid and inspect diaphragm and plunger Visual inspection Remove solenoid and inspect diaphragm and plu...

Страница 6: ...41 3 4 Spud Friction Washer 42 3 4 Spud Sleeve 43 3 4 Spud Escutcheon 44 Setscrew for Control Stop Cover 45 Vandal Resistant Control Stop Cover 46 Stop Cap 47 Adjusting Screw 48 Guide Holder 49 Piston Guide 50 Guide O Ring 51 Piston 52 Piston Seal 53 Stop Body 54 Sweat Solder Adapter 55 Supply Cover Tube Flush Connections and Spud Coupling Kits Product No 3 4 Flush Connection and Spud Coupling Ite...

Отзывы: