13
Ramps
The number of desired program sections can be set here. A Program section (ramp)
consists of: heating rate, end temperature, holding time. 2-6 such program sections can be
programmed. Here, 5 ramps to gradual heating up and 1 ramp for cooling down.
Heating rate 1-5
temperature increase in °C / min (°F / min) to reach the final temperature in the respective
program section. Heating rate 1 is limited to a maximum of 50°C / min. Heating rates 2-5 up
100°C / min.
End temperature 1-5
the respective target temperature of the individual program sections. Adjustable Temperature
range: 250°C - 1650°C.
Holding time 1-5
Duration of the respective program section at the end temperature. Maximum holding time
that can be set for a single ramp: 300 minutes. Maximum holding times to be set across all
ramps: 420 minutes.
Cooling rate
Cooling in °C / min (°F / min) to reach the opening temperature. Adjustable
Cooling range: 1 - 50°C /min
Opening temperature
Indicates the temperature at which the oven opens. The one provided by the factory
The setting range is 250°C - 400°C. Upon request, the limitation of maximum opening
temperature to be set can be changed from 250°C - 1650°C. Please contact your service
partner or us at Zubler Gerätebau GmbH.
2.1.7 Explanation of terms