9650-000820-01 Rev. K
Propaq M Operator’s Guide
13-1
Chapter 13
12-Lead ECG Interpretive Analysis
The Propaq M 12-Lead input is Type CF defibrillator proof.
This chapter describes how to use the Propaq M unit to monitor 12-Lead ECG for adult and
pediatric patients, and how to display 12-Lead ECG Interpretive Analysis information for adult
patients.
Propaq M 12-Lead ECG Monitoring provides simultaneous acquisition and storage of
12-Lead information for adult and pediatric patients, and, when enabled, post-acquisition
Interpretive Analysis for adult patients.
Warning!
•
12-Lead ECG monitoring is intended for the recording of 12-lead ECG signals from
adult and pediatric patients in the supine, resting position -- always ensure that the
patient is kept motionless during 12-lead ECG signal acquisition and analysis. Use of
the device to acquire ECG signals from moving or shaking patients may produce noisy
signals that are difficult to interpret.
•
The 12-Lead Interpretive algorithm's interpretive statements are designed to enhance
the diagnostic process. They are no substitute for the qualified judgment of a properly
trained clinician. As with any diagnostic test, always give consideration to patient
symptoms, history, and other relevant factors.
•
12-Lead Interpretive Analysis is for use with
only
adult patients.
•
It is important to enter each patient’s age and gender prior to performing ECG
analysis using the Inovise 12L Interpretive Algorithm. Providing patient age and
gender will ensure that highest accuracy of ECG analysis is attained. If age is not
provided, a default of 45 years is used. If gender is not provided, the default is male.
•
Excessive body hair or wet, sweaty skin may interfere with electrode adhesion. Remove
the hair and/or moisture from the area where the electrode is to be attached.
•
Remove ECG electrodes from their sealed package immediately prior to use. Using
previously opened or out-of- date electrodes may degrade the ECG signal quality.