Note

For monitoring times exceeding 6 hours, Nafion 
tubing is recommended.

Selecting a Sidestream Cannula 

Select a sidestream cannula based on the patient's size 
and monitoring situation. Nasal and nasal/oral cannulae 
are disposable and single patient use.

Attaching the LoFlo Module Cable and 
Inserting the Sample Cell

Follow these steps:

1. Attach the LoFlo module cable to the CO

connector 

at the back of the M Series unit.

2. Remove the LoFlo sampling cannula or airway 

adapter kit from the package.

3. Insert the LoFlo sample cell into the LoFlo sample 

cell receptacle and ensure that it clicks into place.

4. Ensure that the LoFlo module exhaust tube vents 

gasses away from the module environment. 

5. Turn the selector switch on the M Series to 

MONITOR (ON for AED units).

6. Wait for the CO

module to warm up. 

The unit will display the "WARM UP" message for 
approximately one minute while the module and 
accessory warm to operating temperature. The 
message disappears when the module is ready for 
use. 

Note

Warm up time varies with ambient temperature of 
the module.

Zeroing the LoFlo CO

2

 Module 

The first time a particular LoFlo module is connected to 
the M Series, the unit displays the message “ZERO CO2 
MODULE” following the warm up period.
The unit retains the LoFlo module zero settings during 
and after its use. When the previously used module is 
reconnected to the M Series unit, module zeroing does 
not have to be repeated, as long as no other LoFlo 
module has been connected and zeroed during the 
interim period. 

Note

After the initial zero procedure, the LoFlo module 
may be used during the warm up period. The 
capnogram, EtCO

2

, and respiratory rate will be 

displayed; however, full product specifications are 
not achieved until the warm up period has 
completed.

CAUTION!

Always ensure that the sample cell is 
properly connected to the LoFlo module 
before zeroing.

1. Ensure that the nasal cannula or on-airway adapter is 

not connected to the patient or close to any source of 
CO

(including the patient's -- and your own -- 

exhaled breath and ventilator exhaust valves). 

Pediatric/Infant Airway Adapter Kit

4.0 mm

Pediatric/Infant Airway Adapter Kit 
with Nafion tubing

Cannula Application

Nasal CO

2

 Sampling Cannula, Adult

Nasal CO

2

 

sampling only

Nasal CO

2

 Sampling Cannula, 

Pediatric

Nasal CO

2

 Sampling Cannula, Infant

Oral/Nasal CO

2

 Sampling Cannula, 

Adult

Oral/Nasal CO

2

 

sampling only

Oral/Nasal CO

2

 Sampling Cannula, 

Pediatric

Nasal CO

2

 Sampling with O

2

 Delivery 

Cannula, Adult

Nasal CO

2

 

sampling with 
oxygen delivery

Nasal CO

2

 Sampling with O

2

 Delivery 

Cannula, Pediatric

Oral/Nasal CO

2

 Sampling with O

2

 

Delivery Cannula, Adult

Oral/Nasal CO

2

 

sampling with 
oxygen delivery

Oral/Nasal CO

2

 Sampling with O

2

 

Delivery Cannula, Pediatric

Airway Adapter Kit

ET Tube Diameter

Figure 4

Module Cable

Exhaust Tube

Sample Cell

Sample Cell
Receptacle

LoFlo Module

Содержание M Series

Страница 1: ...9650 0223 01 Rev D TM End Tidal Carbon Dioxide EtCO2...

Страница 2: ...l Corporation CAPNOSTAT is a registered trademark and LoFlo is a trademark of Respironics Novametrix LLC Cidex is a registered trademark of Advanced Sterilization Products a Johnson and Johnson Compan...

Страница 3: ...mum filtration of fluids and contaminants and protecting the system from aspiration of these fluids In both systems the CO2 sensor generates infrared light and beams it through the airway adapter or s...

Страница 4: ...es EtCO2 option Activate oxygen compensation if O2 levels in excess of 60 are introduced Activate N2O compensation if nitrous oxide is introduced into the airway circuit The presence of Desflurane bey...

Страница 5: ...h windows in a vertical NOT a horizontal position This helps keep patient secretions from pooling on the windows Do NOT insert any object other than the sample cell into the sample cell receptacle on...

Страница 6: ...t and property even if ZOLL has been advised of the possibility of such damages THE WARRANTY SET FORTH HEREIN IS EXCLUSIVE AND ZOLL MEDICAL CORPORATION EXPRESSLY DISCLAIMS ALL OTHER WARRANTIES WHETHER...

Страница 7: ...EtCO2 setup as follows Attaching the CAPNOSTAT 3 sensor cable Selecting a mainstream airway adapter Attaching the airway adapter to the CAPNOSTAT 3 sensor Zeroing the mainstream sensor airway adapter...

Страница 8: ...ximately 15 seconds The unit displays the message ZERO DONE upon completion of the zeroing Note Do not attempt zeroing for 20 seconds after removing the adapter from the patient s airway This time all...

Страница 9: ...ow steps a through c else go to step 4 a Verify proper connection of the adapter to the sensor b Verify that the airway adapter windows are clean and dry c If the adapter is properly connected and the...

Страница 10: ...after its use When the previously used module is reconnected to the M Series unit module zeroing does not have to be repeated as long as no other LoFlo module has been connected and zeroed during the...

Страница 11: ...lo module and ensure that it clicks into place 2 If the unit displays either of the following messages take the appropriate action 3 Place the airway adapter assembly at the proximal end of the airway...

Страница 12: ...o module and rubber protective covering can be cleaned and disinfected by wiping with one or more of the following solutions 70 isopropyl alcohol 2 gluteraldehyde 10 bleach solution Use a soft cloth d...

Страница 13: ...ftkey from the physiological monitoring menu cycles the display from the capnogram waveform to the plethysmograph waveform if SpO2 is installed to no second waveform displayed Use the Zoom softkey fro...

Страница 14: ...airway circuit If the concentration of oxygen in the breathing circuit exceeds 60 and nitrous oxide is in use both O2 and N2O should be activated When the Comp softkey is pressed the unit displays the...

Страница 15: ...alarm functions are disabled Background ECG analysis functions continue to operate as described in the AED section of the M Series Operator s Guide The ALARM SUSPEND button can be used to activate dea...

Страница 16: ...ature unit displays WARM UP message 4 Perform a zero procedure if necessary see Zeroing the LoFlo CO2 Module on page 8 5 Breath normally into the adapter 6 Verify that the unit displays EtCO2 readings...

Страница 17: ...3 kPa 0 95 mmHg OFF 0 12 5 OFF 0 12 6 kPa OFF High Respiration Rate Alarm Limit 120 respirations per min 5 150 respirations per min OFF Low Respiration Rate Alarm Limit 5 respirations per min 0 100 re...

Страница 18: ...PU Oral Nasal CO2 sampling only pediatric Nasal CO2 Sampling with O2 Delivery Cannula Adult 8000 0356 SPU Nasal CO2 sampling with O2 delivery adult Nasal CO2 Sampling with O2 Delivery Cannula Pediatri...

Страница 19: ...problem persists return for service CO2 DEVICE NOT READY The zero operation cannot be initiated because The sensor or module is still warming up No sensor or module is attached to the unit Wait for se...

Страница 20: ...g the adapter type Clean airway adapter and reattach it If the problem persists or the adapter type was changed perform a mainstream airway adapter zero as described in Zeroing the Mainstream Sensor A...

Страница 21: ...line is set below 0 mmHg Perform a sidestream module zero as described in Zeroing the LoFlo CO2 Module on page 8 CHECK CO2 LINE Sample line blockage Verify that the sample line is plugged into the mod...

Страница 22: ...han 5 minutes Replace module ZERO CO2 MODULE Module cable connected to M Series unit for the first time Probe drift error detected or negative CO2 detected Perform module zero as described in Zeroing...

Страница 23: ...lt pediatric adapters 50 ms Neonatal adapters 200 ms Respiration Rate Range 0 150 breaths per minute Respiration Rate Accuracy 1 breath per minute Sample Flow Rate N A 50 ml min Compensations Barometr...

Страница 24: ...at maximum energy 360J or 1 5 hours minimum of continuous ECG monitoring or 1 0 hour of continuous ECG monitoring pacing at 60 mA 70 beats min For a new fully charged XL battery pack at 20 C 60 defibr...