background image

5

General Information

Specifications 

ZOLL Base PowerCharger

4x4

General

Indicator Lights

Controls

ZOLL Batteries

Size

4.2 “ H x 13.2 “ W x 12.2 “ L inches

Weight

4.1 kg (9.0 lbs) without batteries.

Equipment Type

IEC Type B

Power 
Requirements

100-120 VAC, 50/60 Hz input.
220-240 VAC, 50/60 Hz input. 

Power 
Consumption

100 watts maximum. 

Design 
Standards

Designed to meet or exceed UL 544, 
EN 60601, and CSA 22.2 standards for 
medical equipment safety.

Recharge Time

QuickCharge

Total recharge of fully depleted battery:

4 hours or less for the PD 4410/Smart/
Smart Ready Battery. 

8 hours or less for all XL Batteries.

AutoTest

AutoTest and total recharge of fully 
depleted battery:

8 hours or less for the PD 4410/Smart/
Smart Ready Battery

15 hours or less for all XL Batteries

Temperature

5 °C to 40 °C (operating)
15°C to 35°C (optimal battery charging)
-40°C to 70°C (storage and transportation)

Humidity

5% to 95% (relative humidity - non 
condensing).

External Fuses

2 Type T, 2A/250V Time Delay fuses 
located above AC mains input. 

POWER

Illuminates when ZOLL Base 
PowerCharger

4x4

 is plugged into live AC 

mains. 

CHARGER ON

Indicates battery is being charged. 
Separate indicator light for each battery 
compartment.

BATT.READY

Illuminates when the charge cycle is 
complete. Separate indicator light for 
each battery compartment.

FAULT

Illuminates when battery capacity test or 
battery compartment test fails. Separate 
indicator light for each battery 
compartment.

TESTING

Illuminates when either a battery test or a 
battery compartment test is in progress. 
Separate indicator light for each battery 
compartment.

ZOLL Base PowerCharger

4x4

 with 

AutoTest only

Illuminates with the 

CHARGER ON 

light 

during AutoTest and recharge of the 
battery

TEST Button

Illuminates TESTING indicator light and 
initiates battery test

 

cycle or battery 

compartment test.
Automatically charges, discharges and 
recharges the battery.
Separate indicator light for each battery 
compartment.

Type

Rechargeable, sealed lead acid:

ZOLL Battery Pack PD4410
ZOLL Smart Ready Battery 
ZOLL Smart Battery
ZOLL XL Smart Ready Battery
ZOLL XL Smart Battery

Weight

PD 4410/Smart/Smart Ready Battery: 1 kg 
(2.2 lbs) 
XL Battery: 1.68 kg (3.7 lbs)

Size

PD 4410/Smart/Smart Ready Battery: 

4 cm H x 7 cm W x 18.9 cm L
(1.6 in. x 2.8 in. x 7.4 in.)

XL Battery: 

9.4 cm H x 7 cm W x 20 cm L
(3.7 in. x 2.8 in. x 7.9 in.) 

Capacity

PD 4410/Smart/Smart Battery: 

2.5 Amp Hour rating for new batteries

XL Battery: 

4.5 Amp Hour rating for new batteries

Nominal Battery 
Voltage

2.0 V/cell; 5 cells.

Temperature

0 °C to 50 °C (operating)
15°C to 35°C (optimal battery charging)
-40°C to 70°C (storage and transportation)

Humidity

5% to 95% (relative humidity - non 
condensing).

Capacity Test

At a minimum, test batteries every 
three (3) months. Refer to ZOLL’s Battery 
Management Program brochure.

Содержание Base Powercharger 4x4

Страница 1: ...ZOLL Base PowerCharger4x4 Operator s Manual Base PowerCharger 4x4 TM CAUTIO N DO NOT BLOCK VENTS OR PLACE ANYTHING ON TOP OF THE VENTS HEAT DAMAGE MAY OCCUR Part Number 9650 0054 Rev G...

Страница 2: ...North America Consult your purchasing agreement for terms and conditions associated with your warranty Outside of North America consult the ZOLL authorized representative In order to maintain this wa...

Страница 3: ...UH DWWHU RPSDUWPHQW 7HVWLQJ 3URFHGXUH DXOW RGHV DQG HILQLWLRQV 4XLFN 5HIHUHQFH KDUW IRU 2 DVH 3RZHU KDUJHU 4x4 ZLWK 4XLFN KDUJH 6HFWLRQ 2SHUDWLRQ XWR7HVW 2SHUDWLQJ RQWUROV DQG QGLFDWRUV 6HWXS DWWHU KD...

Страница 4: ...ZOLL Base PowerCharger 4x4 Operator s Manual ii This page intentionally left blank...

Страница 5: ...er 4x4 with AutoTest automatically tests battery capacity with each battery recharge In addition the ZOLL Base PowerCharger 4x4 with AutoTest illuminates the Batt Ready indicator when the battery is f...

Страница 6: ...HVKRRWLQJ This section describes several routine checks to assure proper unit operation It includes cleaning and maintenance instructions and troubleshooting guides intended for use by non technical p...

Страница 7: ...This warranty does not cover items subject to normal wear and burnout during use including but not limited to lamps fuses batteries patient cables and accessories d The foregoing warranty constitutes...

Страница 8: ...epeated use of partially charged batteries will reduce their capacity and operating life Do not use a battery unless the BATT READY indicator is lit and the FAULT light is off Failure to do so may res...

Страница 9: ...compartment FAULT Illuminates when battery capacity test or battery compartment test fails Separate indicator light for each battery compartment TESTING Illuminates when either a battery test or a bat...

Страница 10: ...order to allow tracking of loan equipment You will be given a Return Claim Number under which to return the ZOLL Base PowerCharger 4x4 Send the unit in its original packaging to ZOLL Medical Corporati...

Страница 11: ...x4 refer to QuickCharge unless otherwise specified If your device does not indicate QuickCharge or AutoTest on the label in the upper right corner refer to the QuickCharge section of this manual for p...

Страница 12: ...1 Illuminates when either a battery capacity test cycle is in progress or a battery compartment test is completed Illuminates with the BATT READY light indicating a usable fully charged battery capab...

Страница 13: ...nsecutive beeps to indicate that the battery is not fully charged and should not be used In addition all four indicator lights will illuminate for four 4 seconds and then go off Repeated use of partia...

Страница 14: ...tor lights will illuminate for four 4 seconds and then go off DWWHU RPSDUWPHQW 7HVWLQJ 3URFHGXUH The ZOLL Base PowerCharger 4x4 has four battery compartments that are able to test and charge batteries...

Страница 15: ...d range Low Constant Current Amount of current during a charge was below the specified range High Constant Current Amount of current during a charge was above the specified range VCC Voltage Internal...

Страница 16: ...e ZOLL Base PowerCharger 4x4 will beep to indicate the battery has been removed and should NOT be used The POWER light should be on as long as the unit is plugged into a live AC source 23 5 7 21 4XLFN...

Страница 17: ...r 4x4 refer to AutoTest unless otherwise specified If your device does not indicate QuickCharge or AutoTest on the label in the upper right corner refer to the QuickCharge section of this manual for p...

Страница 18: ...tment or any other failure occurs There are four FAULT indicator lights Each one corresponds to the battery charging testing compartment directly above 7 67 1 Illuminates when either a battery capacit...

Страница 19: ...cific defect see Fault Codes and Definitions on page 17 and take corrective action Before disposing of the battery test the battery compartment for defective operation see Battery Compartment Testing...

Страница 20: ...ry will cause an audible warning of five 5 consecutive beeps to indicate that the battery is not fully charged and should not be used All four indicator lights will illuminate for four 4 seconds and t...

Страница 21: ...range Low Constant Current Amount of current during a charge was below the specified range High Constant Current Amount of current during a charge was above the specified range VCC Voltage Internal fi...

Страница 22: ...ger 4x4 will beep to indicate the battery has been removed and should NOT be used The POWER light should be on as long as the unit is plugged into a live AC source 23 5 7 21 XWR7HVW 5 5 21 77 5 8 7 7...

Страница 23: ...all battery compartments empty check that all indicator lights except POWER are off 3 Press the TEST button under each compartment to verify that the compartments are fully operational see Battery Com...

Страница 24: ...r power source Check for blown fuse 2 CHARGER ON indicator light does not illuminate when battery is in compartment Check that battery is properly seated in compartment Check that battery contacts are...

Страница 25: ...ent refer to the ZOLL Battery Management Program part number 9650 0019 2 DWWHU 3DFNV ZOLL PD 4410 Smart Smart Ready Battery and XL Battery Packs are five cell assemblies of sealed lead acid batteries...

Страница 26: ...ZOLL Base PowerCharger 4x4 Operator s Manual 22 This page intentionally left blank...

Страница 27: ......

Страница 28: ...336 ZOLL Medical France 189 Rue Nationale 10 78310 Coignieres FRANCE Telephone 33 1 30 05 14 95 Telefax 33 1 30 49 06 93 Middle East Africa Regional Office 14 Persefonis Ekali 14578 Athens Greece Tele...