Page
B-7
User Guide
Electromagnetic Immunity Declaration (EID)
Table B-7
Guidance and Manufacturer’s declaration – Electromagnetic immunity for Platform
The Platform is intended for use in the electromagnetic environment specified below. The customer or user of the
Platform should ensure that it is used in such an environment.
Immunity test
IEC 60601 Test Level
Compliance Level
Electromagnetic
environment – guidance
Electrostatic Discharge
(ESD) IEC 61000-4-2
± 8 kV contact
± 15 kV air
± 8 kV contact
± 15 kV air
Floors should be wood,
concrete, or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be at
least 30%
Electrical fast transient/burst
IEC 61000-4-4
Not Applicable
Not Applicable
No mains power connection
Surge IEC 61000-4-5
Not Applicable
Not Applicable
No mains power connection
Voltage dips, short
interruptions, and voltage
variations on power supply
input lines IEC 61000-4-11
Not Applicable
Not Applicable
No mains power connection
Power frequency (50/60 Hz)
magnetic field. IEC 61000-
4-8
30 A/m
30 A/m
Power frequency magnetic
fields should be at levels
characteristic of typical
location in a typical
commercial or hospital
environment.
Содержание AutoPulse 100
Страница 1: ...P N 12555 001 Rev 9 Resuscitation System Model 100 User Guide...
Страница 8: ...Page viii User Guide This page left intentionally blank...
Страница 38: ...Page 2 16 User Guide This page left intentionally blank...
Страница 66: ...Page 4 6 User Guide This page left intentionally blank...
Страница 72: ...Page 5 6 User Guide This page left intentionally blank...
Страница 74: ...Page A 2 User Guide This page left intentionally blank...