Using the Aspirator User Interface
906-0330-01-01 Rev. D
ZOLL 330 Multifunction Aspirator Operator’s Guide
4-9
:
Choosing a Procedure, Vacuum Level, and Use of Smart Flow
To choose a procedure and set the vacuum level for it, follow these steps:
1. Press the parameter button associated with the procedure (Surgical, Pharyngeal, Tracheal, or
CASS).
2.
Press the
Confirm
button to use the initial vacuum level for the procedure, or change it by
turning the
Dial
clockwise (to increase the vacuum level) or counterclockwise (to decrease
the vacuum level), then press the
Confirm
button.
3.
Turn Smart Flow ON or OFF, by pressing the
Smart Flow
parameter button, then turn the
Dial
clockwise (Smart Flow ON) or counterclockwise (Smart Flow OFF). See the “Smart
Flow” and “Smart Flow Calibration” sections, in this chapter, for more information.
Intermittent Mode Operation
Intermittent Mode allows you to control the following parameters.
•
Vacuum level.
•
ON Time
•
OFF Time
The Intermittent Mode vacuum range, adult and pediatric initial vacuum level settings are:
CASS
The Continuous Aspiration of Subglottic Secretions (CASS) procedure uses
low level of vacuum applied either with a special endotracheal tube (ETT) or
with an adjunct to a standard endotracheal tube to provide suction during
mechanical ventilation. The vacuum removes the secretions that collect
above the cuff of the ETT that are known to increase the risk of ventilator
associated pneumonia (VAP).
Procedure
Vacuum Range
Adult Initial
Vacuum Level
Pediatric Initial
Vacuum Level
Surgical
20 to 550 mm Hg
500 mm Hg
500 mm Hg
Pharyngeal
60 to 400 mm Hg
300 mm Hg
300 mm Hg
Tracheal
60 to 220 mm Hg
100 mm Hg
80 mm Hg
CASS
10 to 40 mm Hg
30 mm Hg
10 mm Hg
Procedure
Description
Control
Vacuum Range
Adult Initial
Vacuum Level
Pediatric Initial
Vacuum Level
Vacuum
10 to 200 mm Hg
90 mm Hg
80 mm Hg
ON Time
5 to 40 seconds
15 seconds
15 seconds
OFF TIme
5 to 40 seconds
40 seconds
40 seconds
Содержание 330
Страница 1: ...ZOLL Model 330 Multifunction AspiratorOperator s Guide 906 0330 01 01 Rev D...
Страница 30: ......
Страница 40: ...Using the Aspirator User Interface 4 6 www zoll com 906 0330 01 01 Rev D...
Страница 66: ......