2.1 Introduction
The instrument may be used under high voltage. In order to avoid injury or death to
operators caused by electric shock or other hazards, please read carefully and fully
understand the content of this chapter before installing, using or repairing this
product.
2.2 Safety Operation
Ensure that the operator who uses this instrument has already read and understand
the operation instructions and safety precautions.
The instrument can only be used under specific environmental conditions. Ensure
that the actual conditions are in accordance with the allowable conditions described
in the "Technical Parameters" section.
Make sure the vent of instrument is not blocked during operations.
Always comply with the explanation in the “Transport and Storage” of chapter 1.
2.3 Usage
Ensure that the voltage of tested signal does not exceed the rated value. The
instrument should only be used to test specified types of signals. For details, please
see the "Technical Parameters" section.
The damage caused by improper use of the instrument is not within the scope of the
warranty.
2.4 Quality Assurance Period
The quality assurance period of this instrument is one year from the date of purchase.
2.5 Protection Limit Values of Accessories
The input of CAT
Ⅱ
(300V) IEC Measurement
Ⅱ
Class can be connected to the
power supply (maximum 300VAC) under overvoltage conditions.
Banana plug: CAT
Ⅱ
1000V/Max. 32A
Testing hook: CAT
Ⅲ
1000V/Max. 10A
Alligator clip: CAT
Ⅱ
300V/Max. 15A
Testing probe: CAT
Ⅲ
1000V/Max. 10A
In order to avoid device damage and electric shock hazard, all above protection limit