Oxygen Monitor SGM7.2 5 Technical Data
5.4 Interface
data
Serial Interface RS-232
Transfer rate .................................. max. 19200 Baud, adjustable
Stop bits ......................................... 1
Data bits ......................................... 8
Parity .............................................. none
Handshake ..................................... without
MD
TXD
RXD
Pin-Configuration Connector SUB-D 9-pol. F:
Table 1: Serial Interface Protocol (CR = Carriage Return)
Command Returned
Data
Description
M2CR
M2x.xxExxCR
Oxygen concentration in ppm
A1CR
A1xxxCR
Cell voltage in mV
A2CR
A2xxxCR
Measuring cell temperature in °C
Table 2: Error Indication
Error code
Description
Error0 Transmission
error
Error1 Warm
up
Error2
Cell temperature to low
Error3 Thermocouple
defect
Error6 System
error
Analog output
Current signal ......................... 0/4 - 20 mA, potential free, range adjustable,
apparent ohmic resistance max. 500 Ohm
Alternative voltage signal ....... 0/2 - 10V (on request), potential free,
range
adjustable
Table 3: Properties limit value switch
Load
Resistive load (cos
ϕ
= 1)
Max. switching voltage
125 V AC, 60 V DC
Max. switching current
1A
Max. Switching power
62.5 VA, 30 W
Min. permissible load
1 mA at 5 V DC
HB_SGM72_Rehm_eng.DOC
13