Enroll the patient.
Online: www.ZioSuite.com
Phone:
1
.
888
.
693
.
2401
Write in Name, Physician, Start Date
and Time on cover of
Patient Instructions
& Button Press Log
.
Determine position:
Patient should stand with arms
relaxed at side.
Skin should be free of perspiration.
1. Plan & position
S
tart
G
uide
Determine placement area without
removing backing.
Place on flattest part of upper left chest:
• 1 finger width below collarbone,
centered over left pectoral muscle
• Edge of the Zio XT Patch next to sternum
• Angle so arrow on top label
points
upward
• Avoid armpit and breast tissue
Prepare a large area:
Shave
area.
Abrade
skin, applying pressure for 40
broad strokes—10 times in the direction
2. Prep skin
of each arrow in 2a & 2b (important for
adhesion and ECG signal quality).
Clean
skin thoroughly with all alcohol
pads. Let dry for 1 minute.
1
2a
2b
3
4c
4b
4a
Hold device in the center and remove
clear backings. Keep top label on. Do not
touch adhesive.
3. Apply patch
Place Zio XT Patch in the recommended
position.
Press firmly across the entire device for
2 minutes (see image 3).
Peel off the 2 parts of the top label
(see image 4a).
Press firmly across the entire device for
2 minutes, working adhesive into the skin.
Emphasize pressing on the dotted areas
shown in figure 4b.
4. Finish & activate
Firmly press the ‘Zio’ button and release
(see image 4c). The
green
light will flash
5 times indicating that the monitoring
has started.
Help patient get familiar with the location
of the button.
Ask your patients to download the
MyZio™ symptom tracking app.
5. Download & register