background image

Enroll the patient. 
Online: www.ZioSuite.com 
Phone: 

1

.

888

.

693

.

2401

Write in Name, Physician, Start Date  
and Time on cover of 

Patient Instructions  

& Button Press Log

.

Determine position:

 

Patient should stand with arms  
relaxed at side.

Skin should be free of perspiration.

1. Plan & position

S

tart 

G

uide

Determine placement area without 
removing backing.

Place on flattest part of upper left chest:

•  1 finger width below collarbone,    
  centered over left pectoral muscle
•  Edge of the Zio XT Patch next to sternum
•  Angle so arrow on top label  
 points 

upward

•  Avoid armpit and breast tissue

Prepare a large area:

Shave 

area.

Abrade

 skin, applying pressure for 40 

broad strokes—10 times in the direction  

2. Prep skin

of each arrow in 2a & 2b (important for 
adhesion and ECG signal quality).

Clean

 skin thoroughly with all alcohol 

pads. Let dry for 1 minute.

1

2a

2b

3

4c

4b

4a

Hold device in the center and remove  
clear backings. Keep top label on. Do not 
touch adhesive.

3. Apply patch

Place Zio XT Patch in the recommended 
position.

Press firmly across the entire device for  
2 minutes (see image 3).

Peel off the 2 parts of the top label  
(see image 4a).

Press firmly across the entire device for  
2 minutes, working adhesive into the skin. 
Emphasize pressing on the dotted areas  
shown in figure 4b.

4. Finish & activate

Firmly press the ‘Zio’ button and release 
(see image 4c). The 

green

 light will flash  

5 times indicating that the monitoring  
has started.

Help patient get familiar with the location 
of the button.

Ask your patients to download the  
MyZio™ symptom tracking app.

5. Download & register

Отзывы: