Troubleshooting
Problem
Cause
Solution
No power (the Instrument and
Screen remain blank when the
power is turned on)
1.
AC power cord is not
connected.
2.
The main switch on the left
side below the electric
control box is not switched
on.
1.
Check AC power cord
connections at both ends.
Use the correct cord.
2.
Check
t
he main switch,
turn it on.
The self-testing program does
not run (LCM screen turns on
when the power is on)
1.
The packing materials are
not removed.
2.
Technical problem.
1.
Turn off the power and
remove the packing
materials.
2.
Contact your local
representative or agent.
Protocol stops after an initial
start
1.
Consumables are loaded
incorrectly.
2.
Motion sensors are
inactive.
1.
Turn off the power and
turn it back on to stop the
program. The system will
go back to initial start. Re-
load the consumables
according to the
instructions in this manual.
Note: You cannot resume the
protocol after stopping it, you
may lose your samples.
2.
Turn off the power and
remove all samples and
consumables. Contact
your local representatives.