10. Maintenance
If you wish to discard the existing product after purchasing a new product from Remed Co.,
Ltd., we will deal with it free of charge when the new product is delivered. This device and
its components must not be disposed of with household waste and must be disposed of
according to professional disposal procedures. Please proceed according to each country's
disposal procedure. If necessary, please contact our customer service center.
10.1 Cleaning procedure
ZTone and components can be kept clean in a number of ways. Use the methods
recommended below to avoid damage or contamination of the instrument. Before
cleaning the device, be sure to turn off the power of the device. Cleaning the inside of
the device is only permitted by persons authorized by us.
❏
Periodically clean the exterior of the instrument, transducer, and touch screen with a soft
cloth moistened with alcohol, and do not use abrasives, lacquers, thinners, ethylene, or
oxides. These substances may cause permanent damage to the product.
❏
Do not immerse any part of the device in liquids or detergents. Also, do not allow any
liquids to enter the device or components.
❏
At the end of single person use, clean the transducer surface with alcohol and a soft
cloth.
❏
If you want to use a method or disinfectant other than those mentioned above, the device
may be damaged, so be sure to contact us before using it. Please note that if the device
is damaged by using an unauthorized substance, the warranty is not guaranteed even
during the warranty period.
❏
We recommend cleaning once every 1-2 weeks.
10.2 Routine inspection of equipment
❏
The covering of power line of equipment, transducer connecting line, etc. shall not be
peeled off and internal lines shall not be exposed, and shall not be damaged by impact
from outside.
❏
There shall be no trace of oil leakage from transducer.
❏
Clean the outside of equipment so that there is no foreign material.
❏
The button for equipment operation, etc. must not shake.
❏
The various parts attached to equipment must not shake.
❏
If any of the above occur, contact customer service for help.