Checking for malfunctions
Serial
no.
Error
Reason
Measures
1
The switch key lights up
but the LED for the power indi-
cator does not
The battery loses power, the
adapter works irregularly
Check the plug
between the cable and power
2
The LED for the power indicator
lights up, but the LED for the
image does not
The interval time is too
short for a restart
Restart after 1 minute
3
The LED displays a character
menu but no scan image
1.
Error when starting up
power, with the gain or
STC control
2.
No connection to the
probe, or the probe
connection is not correct
3.
The device is in freeze-
frame
mode:
1.
Control power start up, gain
or STC control
2.
Ensure that the connection is
correct.
3.
Exit standby mode
4
Abnormal image
1.
Examination mode errors
2.
Image processing set-
tings errors
1.
Check whether the examina-
tion mode is suitable or not
2.
Adjusting the image pro-
cessing
Setting as default
5
The probe is not working
properly
1. The plug is loose
2. The internal circuit pro-
tects
1. Pull the probe out and rein-
sert it
2. Restart
6
No menu for the birth prepara-
tion calculation package
Do not select the obstetrics
application before
scanning.
Select the obstetrics application
7
The PRINTER button does not
work
1.
The connected printer is
approved
2.
The printer is not turned
on
3.
The printer is not correctly
connected
1.
Change the approved printer
2.
Turn the printer on
3.
Connect the printer again