110
CAUTION:
Do not immerse the probe connector in any liquid
Do not exceed the insertion depth of the intra-cavitary probe. It is essential to avoid leaving the probe in
the liquid for more than twelve hours. Use only the appropriate test solution.
Storage:
Please place the probe to a clean and dry environment, and avoid direct sunshine.
Maintain a temperature range of between -10 and 50º C in the environment where the probe is stored, and
avoid any high pressure or vacuum environments.
Always take care when handling the probe and avoid damage.
The probe should always be stored in its case during transportation or when not in use.
8.3
Safety check
The system is listed in neither Appendix 1 nor Appendix 2 of the MPBetreibV (German Medical Products Operator Or-
dinance).
In Germany, the DGUV Regulation 3 (Electrical Systems and Equipment) in its current version must also be observed.
NOTE:
This note applies to operation of the system in Germany. Where applicable, comply with any divergent national regula-
tions in your country.
To make sure that the device is functioning normally, it is recommended to prepare a maintenance plan to check the
device regularly. If the device functions abnormally in any way, its use must be stopped immediately, and the sales
office, customer service department or representative of the manufacture must be contacted.
If no image is displayed or the menu is displayed without an image, please check for possible malfunctions using the
following table. If the error cannot be resolved, please contact the sales office, customer service department or rep-
resentative of the manufacturer.