Maintenance
5.1 Cleaning
5
Page 30
Cleaning
The handpiece and the tips directly touch the skin of the person being treated.
Therefore, it is necessary to clean and disinfect it after each use. For cleaning and
disinfection, first use a soft, damp cloth, followed by a dry cloth.
1. Roll handpiece
•
Clean the suction opening with a cotton swab or a connection cleaning brush.
Be especially careful and avoid blockages.
2. Multipolar
handpiece
•
Wipe the skin contact area first with a damp cloth and then with a soft dry
cloth.
•
Clean the suction opening with a cotton swab or a connection cleaning brush.
3. Thermal handpiece
•
Wipe the ceramic electrodes first with a damp cloth and then with a soft, dry
cloth.
4. Device and screen
•
Clean the outside of the system and LCD main display with a soft cloth. A
mild cleaning agent can be used, but be careful that no liquid gets into the
device.
•
If heavily soiled or if there are deposits, clean the device several times
carefully as described above. Avoid any pressure when wiping.
5. Air filter
1.
Unscrew the screws at the back in order to reach the air filter.
2.
Disconnect the air filter from the main unit by pressing the black lever in the
direction of this symbol ( ) and turn the air filter to the left.
3.
Remove the oil, clean the filter using a mild soap, and dry it completely.
4.
Secure the bottle in the device by turning the black lever in the direction of
this symbol (
) and then turning the air filter to the right.