Safety / Maintenance / Service
12.2 Service
12
Page 22
Service
Activating the “Service” button in the configuration menu opens the “User”
service screen.
Defrost
We recommend daily defrosting (in the evening) if Cryo 7 is operated for
several hours, since ice may have formed around the evaporator.
Please empty the defrost container before the defrosting process.
Activating the “Start” button in the “Defrost” display field starts the
defrosting
process.
While the defrosting process is running, no screen entries or therapy is
possible.
After the end of the defrosting process, the defrost container must be emptied.
Capacity of the
defrost container
The defrost container has a capacity of approx. 1.5 litres.
If the defrost container has reached a certain fill level, an acoustic signal
sounds. In addition, the “Defrost container fill level” icon in the status bar
switches to the “Empty defrost container” icon.
Maximum capacity
reached
Cryo 7 also displays a message in the therapy screen to empty the defrost
container.
Emptying
To remove the defrost container, open the maintenance door.
Remove the defrost container from the hatch and take the cover off the top of
the defrost container.
After you have emptied the defrost container, ensure that the cover is
positioned correctly on the defrost container.
Insertion
Insert the defrost container as follows:
•
Opening is on top and facing the device.
•
Because of the defrost valve, there is slight resistance when inserting
the defrost container in the device; this can be overcome by gentle
pressure.
•
Push the defrost container all the way into the device.
•
Close the maintenance door.
Note
The defrost container must always be in the device, except during emptying
and cleaning.
S01 / S02
The S01 and S02 maintenance programs simulate long-term operation and
document the proper condition of Cryo 7.
Any faults are recorded
Note
Before starting the maintenance program, the defrost container must be
emptied.
Содержание Cryo 7
Страница 1: ...Instructions for Use Cryo 7 EN ...