Zimmer Biomet Tricera Скачать руководство пользователя страница 12

 

DCO-1318 

 

Effective Date: 01-SEP-2021

 

LBL 6028 Rev. G 

Page 12 of 23 

 

7.

 

To perform soft tissue resection, bone burring, or ablation, position 

the tip of the device adjacent to the tissue and press the activate 

button/pedal.  

8.

 

If bleeding occurs, position the electrode adjacent to the bleeding 

tissue and activate the blue pedal on the footwitch or blue button on 

the Handpiece.  

9.

 

The handpiece joystick can be used to adjust rotational speed, set 

pressure, suction level, and/or RF power level depending on the 

attached device, mode, and joystick setting. See section 15.2-15.3 

for more detail. 

10.

 

To change the mode, the left and right black footswitch corner 

buttons can be pressed, or the mode button on the handswitch can 

be used. 

11.

 

To change the coagulation power level, press the coag power level 

button next to the settings button in the bottom left of controller GUI.  

12.

 

To activate suction at any time, press the black suction pedal on the 

right-hand side of the footswitch or hold the mode button on the 

handpiece. 

12. 

DISASSEMBLY 

1.

 

Remove the Tricera Device from the cavity. 

2.

 

Applies to Veriflow FMA set R-10003 only: Before the removal of 

the arthroscope from the cavity, turn off saline pump by pressing the 

flow button on the footswitch or the stop button on the GUI screen.  

3.

 

Remove the red suction tubing connector from the Tricera 

Handpiece. 

4.

 

Remove the Tricera Device from the shaver handpiece.  Dispose of 

the Tricera Device following standard hospital practices. 

 

Steps 5-7 apply to Veriflow FMA set R-10003 only  

5.

 

Disconnect the inflow tube from the Arthroscope/Sheath. 

6.

 

Remove the saline bag from the saline pole and bring to a height 

lower than the controller. 

7.

 

Disconnect the saline spikes. 

8.

 

Press the unlock button on the controller screen to remove the FMA 

cassette.  The FMA cassette can now be disposed following 

standard hospital practices. 

9.

 

Disconnect the Tricera Handpiece from the Tricera Controller. 

10.

 

Follow the cleaning and sterilization procedures for the Tricera 

Handpiece per its instructions for use. 

13. 

CLEANING 

13.1.

 

Cleaning Precautions

 

1.

 

Follow universal precautions for protective apparel when handling 

and cleaning contaminated instruments. 

2.

 

Clean instruments within 30 minutes after use to minimize the 

potential of blood and debris drying prior to cleaning. 

3.

 

After use, Tricera Devices may be a potential biohazard and should 

be handled in accordance with acceptable medical practice and 

applicable local and national requirements. 

4.

 

Dispose of all non-reusable tubing sets, irrigation tips and Tricera 

Devices properly after use. 

5.

 

DO NOT lubricate any handpieces. 

6.

 

DO NOT immerse the controller, footswitch or attachments. 

7.

 

Clean the controller, footswitch and footswitch cable per the 

Cleaning Instructions in Section 13.2. 

8.

 

Clean the handpieces per the Cleaning Instructions in the 

Instructions for use for the Tricera Handpiece. 

9.

 

Never clean handpieces with bleach, chlorine-based detergents, 

liquid or chemical disinfectants, or any products containing sodium 

hydroxide.  They will degrade the coatings. 

10.

 

Never clean equipment in an ultrasonic cleaner or a combination 

washer/sterilizer. 

11.

 

For aluminum surfaces, a neutral-pH agent should be used.  To 

prevent corrosion, avoid contact with strong alkaline solutions (pH 

over 10.5) or agents containing iodine or chlorine. 

13.2.

 

Tricera Controller and Footswitch Cleaning Instructions

 

Follow this procedure after each operation to clean the Controller and 

Footswitch: 
1.

 

Disconnect the Controller from the electrical source.   

2.

 

Wipe the controller, footswitch and footswitch cable with a clean, 

soft cloth dampened with a mild, pH-balanced detergent.  Prolonged 

exposure to any corrosive solvents or disinfectants should be 

avoided. 
a.

 

Thoroughly clean the tubing cassette roller area on the 

controller.  Ensure the rollers are free of all debris. 

b.

 

Be sure to wipe the underside of the footswitch to ensure no 

particles have attached themselves. 

3.

 

Wipe equipment again with distilled or sterilized water to prevent 

metal discoloration. 

13.3.

 

Tricera Handpiece Cleaning Instructions 

See the Instructions for use for the Tricera Handpiece 

14. 

STORAGE 

14.1.

 

Tricera Controller (See Appendix A) 
14.1.1.

 

Fluid Management Accessories and Tricera Devices 

The unused Tricera Devices and Fluid Management 

Accessories should be stored at room temperature, 

away from moisture and direct heat.

 

14.2.

 

Tricera Handpiece 

The Tricera Handpiece should be stored at room temperature, 

away from moisture and direct heat. 

15. 

TRICERA CONTROLLER SETTINGS, 

MAINTENANCE, TROUBLESHOOTING AND 

REPAIR 

15.1.

 

Adjusting Volume 

Adjusting the volume on the controller is accessed through the options 

menu: 

 

 

Figure 26 

Options Menu 

Содержание Tricera

Страница 1: ...15 Appendix C Abbreviations 15 Appendix D Tones 15 Appendix E Troubleshooting 16 Appendix F Essential Performance Power Curve 18 Appendix G EMC Tables 19 Appendix H User Profiles 21 1 SYMBOLS USED ON...

Страница 2: ...on during diagnostic and operative arthroscopic procedures The Shaver Blade RF Probe of the Tricera Arthroscopic System provides abrasion resection debridement and removal of bone through its shaver b...

Страница 3: ...hin the joint The inflow stopcock should be completely open or completely closed WARNING Keep the pump dry If liquid has accidentally leaked into the pump from the Fluid Management Accessories cassett...

Страница 4: ...ble devices such as cardiac pacemakers If the patient has an implantable cardioverter defibrillator ICD contact the ICD manufacturer for instructions before performing ablation coagulation or other el...

Страница 5: ...ay result in damage to the tip of the devices 7 ENVIRONMENTAL PROTECTION Observe all national waste management regulations DO NOT dispose of Waste Electrical and Electronic Equipment WEEE as unsorted...

Страница 6: ...6 Receptacle reserved for future functionality 7 Power ON OFF Switch and Power ON LED 8 Pressure Sensors Figure 2 Controller Rear 9 Equipotential Lug 10 Power Entry Module 11 Fuse Drawer 12 USB Acces...

Страница 7: ...nt 6 Inflow to Patient 7 Inflow Pressure Diaphragm 8 Inflow Peristaltic Pump Tube 1 TPE Pump Tubing 2 Barb Lock 3 PVC Waste Tuing 4 Outflow Lumen 5 Barb Connector 6 Cassette Magnet Figure 5 Veriflow T...

Страница 8: ...Date 01 SEP 2021 LBL 6028 Rev G Page 8 of 23 Figure 7 Tricera Handpiece LCD and Buttons 1 Activate 2 Coagulation 3 Joystick 4 Mode 9 5 Shaver Blades 1 Coupler 2 Hub 3 Shaft 4 Tip Figure 8 Tricera Devi...

Страница 9: ...x H for user profile details 5 The Zimmer Biomet logo and self test progress will appear on the screen followed by the images shown in Figure 11 and 12 Figure 11 Figure 12 6 Open the FMA Cassette pack...

Страница 10: ...e end of the Tricera handpiece as shown Figure 18 NOTE steps 14 17 apply to Veriflow FMA set R 10003 only 14 Select the joint for the procedure by touching the joint selection touchscreen or using the...

Страница 11: ...at any time during the procedure 2 The Tricera Controller will set the following default pressure set point based on the initial joint selection Knee 45mmHg Shoulder 50mmHg Hip 50mmHg Small Joint 35m...

Страница 12: ...e the potential of blood and debris drying prior to cleaning 3 After use Tricera Devices may be a potential biohazard and should be handled in accordance with acceptable medical practice and applicabl...

Страница 13: ...5 4 Allow 0 Suction YES and NO When this setting is set to Yes the suction level for Oscillate Bone Cutting Bone Burring Forward and Reverse all non RF modes shall be adjustable between 0 and 100 When...

Страница 14: ...of merchantability and fitness for a particular purpose and all other obligations and liabilities on the part of Zimmer Biomet Zimmer Biomet neither assumes nor authorizes any other person to assume f...

Страница 15: ...Pressure 878 to 1082 cmH2O 86 to 106 kPa IV Transport and Storage Requirements Temperature 18 C to 60 C Relative Humidity 15 to 85 non condensing Atmospheric Pressure 510 to 1082 cmH2O 50 to 106 kPa A...

Страница 16: ...tomer service E11 E16 Inflow turns off High pressure detected on the inflow line Foreign object between diaphragm and pressure sensors Controller malfunction Remove any obstructions from inflow line o...

Страница 17: ...t the footswitch is plugged into the gray port on the Controller Ensure the Handpiece is securely plugged into the connector Replace Tricera Device N A N A N A F12 Handpiece Cannot Position Window Pro...

Страница 18: ...ntentional activation of RF output no unintentional activation of pump motors and correct pressure indication within one indicator bar II Power Curve Full and Half Output Power versus Load Resistance...

Страница 19: ...ootswitch 10 ft Hospital Grade Power Cord Table 1 Electromagnetic Emissions Statement Guidance and Manufacturer s Declaration Electromagnetic Emissions The Tricera System is intended for use in the el...

Страница 20: ...lity should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations on power supply input lines IEC 61000 4 11 5 UT 95 dip in UT for 0 5 cycle 4...

Страница 21: ...3 79 3 79 7 27 100 12 0 12 0 23 0 For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be estimated using the equation applicable to...

Страница 22: ...3000 6000 Forward RPM Reverse Default 1000 6000 6000 Reverse RPM Bone Cutting Default 3000 15000 15000 Bone Cutting RPM Bone Burring Default 3000 15000 15000 Bone Burring RPM RF Ablation Default 1 10...

Страница 23: ...handpiece shall continue running when the button is released When a button is pressed again the motor shall stop running In RF modes or when Handpiece Button Latch non RF modes is Off activation shall...

Отзывы: