GPS

®

III System

Processing Instructions

Содержание GPS III

Страница 1: ...GPS III System Processing Instructions ...

Страница 2: ...o 60 ml syringe Withdraw 6 ml of ACD A citrate anticoagulant Mini GPS III System Withdraw 3 ml of ACD A 54 ml of blood is drawn from the patient according to facility protocol Mini GPS III System 27 ml of blood is drawn from the patient according to facility protocol ...

Страница 3: ...GPS III System Step 2 Load Blood Unscrew cap on center port No 1 and discard green packaging post Slowly load blood into center port ...

Страница 4: ...d Place tube into centrifuge Mini GPS III System Purple buckets 7433 must be inserted into the centrifuge Fill GPS III counterbalance 800 0508 with 60 ml of sterile saline and place into opposite side of centrifuge Mini GPS III System Fill GPS III mini counterbalance 800 0505 with 30 ml of sterile saline and place into opposite side of centrifuge ...

Страница 5: ... the lid latch clockwise Latched indicator will illuminate Set speed to 3200 RPM and time to 15 minutes Press green button to start spin Once spin is complete press red button to illuminate the Unlocked indicator Twist latch counterclockwise to open lid ...

Страница 6: ...PS III System Step 5 PPP Extraction To extract the platelet poor plasma PPP remove yellow cap on side port No 2 and connect the 30 ml syringe Slowly tilt the tube while withdrawing PPP Replace yellow cap ...

Страница 7: ...thdraw 2 ml of PRP Mini GPS III System Withdraw 1 ml of PRP With 10 ml syringe attached suspend the platelets by shaking the tube for 30 seconds If the pellet between the buoys is not suspended completely shake tube vigorously Extract the remaining platelet rich plasma PRP contents into the attached 10 ml syringe ...

Страница 8: ...Plasmax Plasma Concentration System Processing Instructions ...

Страница 9: ...tely saturated with plasma Step 1 Load Plasma Plasmax Plus Unscrew yellow cap on port No 1 of plasma concentrator and fill with 25 ml of PPP Replace cap Remove and discard protective cover on tethered white cap Screw white cap onto port No 1 Plasmax Regular Fill with 15 ml of PPP Note Refer to GPS III instructions for how to process PRP and PPP ...

Страница 10: ...ce 800 0512 They should be placed on opposite sides in the centrifuge Spin for 2 minutes at 2000 RPM Plasmax Regular Use Plasmax Counterbalance 800 0510 Plasmax Plus Unscrew red cap on port No 2 Connectsterile10mlsyringeonportNo 2andextract approximately 10 ml of concentrated plasma Plasmax Regular Extract approximately 6 ml of concentrated plasma ...

Страница 11: ...met Biologics sales force and health care providers It is not to be redistributed duplicated or disclosed without the express written consent of Biomet For product information including indications contraindications warnings precautions and potential adverse effects see the Biomet Biologics website at www biometbiologics com 2018 Zimmer Biomet 1792 1 US en REV0118 ...